background image

3

The status word is transmitted in the following format: 

1 = Motion

0 = No Motion

1 = Over Capacity

0 = Not Over Capacity

1 = Below Zero

0 = Zero or Above

1 = Center of Zero

0 = Not at Center of Zero

Pounds/Ounces Mode Selected (Both Scale Models)

ASCII

SO

(%0E) =>

Enable Continuous Output

Disable On-Demand Output

ASCII

SI

(%0F) =>

Disable Continuous Output

Enable On-Demand Output

ASCII

~

(%7E) =>

Transmit Weight in Pounds/Ounces Mode

Data Format:  <STX>bPPPbLBb00.0bOZbXCC<ETX>

where:

<STX>

= ASCII STX

b

= ASCII Space

PPP

= Weight in Pounds

LB

= Text Indicating Pounds

00.0

= Weight in Ounces

OZ

= Text Indicating Ounce

X

= Status Character

M = Motion
B = Below Zero
C = Over Capacity

CC

= Checksum Byte

Checksum Byte - All hexadecimal digits up to but not including the checksum are exclusive-or’d with
each other to form a checksum byte. The byte is further split into two nibbles with the high nibble first.
Each nibble is or’d with an ASCII 0 (%30) to produce the final ASCII checksum value.

ASCII EOT (%04) =>Version Request Command

Causes the scale to transmit scale software revision level in the
following format:  <STX>VVVCC<ETX>

where:

<STX>

= ASCII STX

VVV

= 3 Digit Revision Level

CC

= Checksum Digits

<ETX>

= ASCII ETX

The checksum digits (CC) are formed by exclusive or-ing all three version number digits (VVV) to
produce a checksum byte. This byte is then split into two nibbles with the high nibble transmitted first.
Each nibble is or’d with an ASCII 0 (%30) to form the final ASCII checksum representation.

MAINTENANCE

There are no user replaceable or serviceable parts contained within your AS-300 series scale. Do not
disassemble the scale. Maintenance should be limited to cleaning the exterior of the scale enclosure
with a damp cloth.

CAUTION! Make certain that you have turned the power switch to the OFF position and
removed the power cord from the wall outlet before cleaning the scale.

Should you ever need to replace the fuse, make certain that only fuses of the same type and rating
are used to ensure continued electrical protection. If you have any questions concerning service, you
should contact your scale service representative.

0

1

1

0

X X X X
^

^

^

^

Содержание AS-300D Series

Страница 1: ...AS 300D Series Smart Bench Scales Owner s Manual T DET C O E A Division of Cardinal Scale Manufacturing Co PO BOX 151 WEBB CITY MO 64870 417 673 4631 8527 M214 O1 Rev B 01 03 Printed in USA ...

Страница 2: ... editorial or pictorial content in any manner is prohibited No patent liability is assumed with respect to the use of the information contained herein While every precaution has been taken in the preparation of this book we assume no responsibility for errors or omissions Neither is any liability assumed for damages resulting from the use of the information contained herein All instructions and di...

Страница 3: ... check to make certain that the scale is level If the level indicator does not show a level indication the mounting feet should be adjusted to obtain a level condition To adjust a mounting foot first loosen the lock nut then turn the mounting foot in the required direction to raise or lower that corner of the scale Once a level condition has been achieved secure the mounting foot settings by tight...

Страница 4: ...ed to the scale Pounds Only Mode Selected Optional AS 350D only W Standard Resolution Request to Transmit Scale Weight This command can only be used when the scale mode has been set to the Pounds Only Mode Scale Output Format STX Snnn nn CR where STX ASCII STX nnn nn scale weight in pounds with decimal point CR Carriage Return If the scale weight is below zero over capacity in motion or at center ...

Страница 5: ...OT 04 Version Request Command Causes the scale to transmit scale software revision level in the following format STX VVVCC ETX where STX ASCII STX VVV 3 Digit Revision Level CC Checksum Digits ETX ASCII ETX The checksum digits CC are formed by exclusive or ing all three version number digits VVV to produce a checksum byte This byte is then split into two nibbles with the high nibble transmitted fi...

Страница 6: ... 8527 C099 1A Serial Cable 25 1 8527 B205 08 Load Cell Spacer 26 1 8527 D201 08 Base Assembly 27 1 8527 D210 08 Top Cover 28 1 8527 C211 08 Display Overlay 29 1 6670 0003 RFI Filter Assy 30 1 6610 4046 Fuse Holder 31 1 6610 4019 1 4 AMP Slow Blow Fuse 32 1 8527 B122 0A AC Cable Assembly 33 1 6980 0008 Strain Relief Bshng 34 4 6021 0448 4 40 X 75 Flat Head Screw 35 2 8527 B109 0A Cable Assembly 36 ...

Страница 7: ...5 15 14 17 16 18 31 30 37 38 43 29 32 9 10 35 6 1 25 36 19 39 34 28 21 26 23 5 9 8 40 17 16 20 50 49 2 27 18 3 48 22 24 PART IDENTIFICATION Figure No 2 ...

Страница 8: ...U N A M E D S I G U N E D F A C T R E D I N U S A U N A M E D S I G U N E D F A C T R E D I N U S A ...

Отзывы: