Model DM-100
DM-100 Instruction Manual
Rev. 3.5
Page 4 of 41
1.3
Model 100 Standard Terminal Board (Optional)
If the sensor is ordered with an explosion proof condulet/junction box, the sensor will come with the Model
100 Standard Terminal Board mounted in the J-Box (Figure 7). This terminal board affords the user easy plug-
in connectors for use in wiring the unit during installation, and convenient test points for measuring the
milliamp (mA) output.
R
ED
G
RE
E
N
B
L
UE
WH
ITE
V+
mA
Customer
Wiring
Sensor Wiring
Figure 7 DM Standard Terminal Board
The standard terminal board provides an easy Digital Volt Meter (DVM) connection for reading the mA output
from the sensor.
Two test points (TP1 and TP2) provide a connection across a 10Ω resistor that develops a
10mV voltage drop per 1mA output. Thus a reading of 40mV on a DVM equals 4mA of current. This is an
alternative method to reading the mA output of the sensor for assemblies that do not include the DM-100
Series Display option. Refer to Section 2.6, Field Wiring for more information about the setup of the DM-100
Terminal Board.
1.4
DM-100 Display Terminal Board (Optional)
If the unit is ordered with the optional display or a battery pack (wireless configurations) the unit comes with
the DM-100 Display Terminal Board mounted in the J-Box (Figure 8).
J6
J7
J4
J1
J5
J8
L9
J2
AI
N1
W
/BK
W
/BN
W
/BU
W
/G
N
W/
V
WIRELESS
AI
N2
MA
G
ND
PWR
POWER IN
( SOLAR)
WIRELESS
PROGRAM
B
A
G
ND
24V
MODBUS OUT
( WIRELESS)
24V
G
ND
A
B
MODBUS IN
24V
G
ND
A
B
SENSOR
PWR
G
ND
SCL
SDA
SP1
SP2
DISPLAY
JP1
SW2
SW1
TERM
PROGRAM
POWER
mA
R
ED
B
LA
C
K
B
LU
E
W
H
ITE
Customer Wiring
Figure 8 Model 100 Display Terminal Board