• The Replaceable Emitter Cassettes can then be
replaced with a spare cassette or removed to
another area for cleaning.
• Cleaning of the Replaceable Emitter Cassette can
be accomplished with a stiff brush or lint-free swab
and isopropyl alcohol or other residue-free cleaning
solvent.
• The emitter Replaceable Emitter Cassettes or the
individual removable tungsten emitters can also be
cleaned in an ultrasonic cleaning tank.
Calibration
See figure 4.
We recommend re-calibration every 12 months,
maximum. However, most users will often choose a
much shorter interval, especially if the product being
worked on is very sensitive. On very sensitive
product, we recommend checking the calibration any
time an emitter cassette is removed and replaced.
Warning! Do not attempt to calibrate this unit
without proper equipment such as a charged plate
analyzer. Adjustments to the calibration controls
should be performed by authorized trained
personnel only.
Figure 5. Balance under < ±10 volts. When set under
lab conditions, 5 volts is attainable.
TB-2087
Page 6 of 8
Figure 4. Calibrating C/E Ionizer
Fan 4
Fan 3
Fan 2
Fan 1
< 4 Sec.
decay
< 4 Sec.
decay
< 4 Sec.
decay
< 4 Sec.
decay
48 in.
Unit 60479
Fan 2
Fan 1
< 4 Sec.
decay
< 4 Sec.
decay
Unit 60478
24 in.