background image

DESCO WEST 

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178

DESCO EAST 

- One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Website: 

Desco.com

TB-3062

 Page 5 of 8

© 2013 DESCO INDUSTRIES, INC.

Employee Owned

Frequency of Compliance Verification Testing

NOTE: “The frequency of periodic testing is normally 

specified in corporate operating procedures. …The frequency 

of testing is driven by the amount of risk exposure that can 

occur between tests. For, example, what is the quantity of 

product handled between test periods?” (See ESD Handbook 

ESD TR20.20)  
Per ANNEX A of Compliance Verification ESD TR53 “Test 

Frequency, The objective of the periodic test procedures 

listed in this document is to identify if significant changes in 

ESD equipment and materials performance have occurred 

over time. Test frequency limits are not listed in this 

document, as each user will need to develop their own set 

of test frequencies based on the critical nature of those ESD 

sensitive items handled and the risk of failure for the ESD 

protective equipment and materials.”
Addition information includes: 

•  Worksurface

 - at least quarterly (see ESD TR20.20 

section 5.3.1.13 Periodic Tests) 

•  Footwear

 - “Incoming inspection on a lot sampling basis 

should be performed for all static control footwear.” (see 

ESD TR20.20 section 5.3.3.4 Testing) 

•  Floor

 - “The types of monitoring and type of equipment 

are considerations. In some cases, a simple electrical 

resistance test with a megohmmeter may suffice. In 

others, a static charge generation test may be required.” 

(see ESD TR20.20 section 5.3.4.13 Performance 

Monitoring) 

•  Seating

 - “The recommended electrical resistance 

range for seating is less than 1 x 10

9

 ohms as tested in 

accordance with ANSI/ESD STM 12.1. This value should 

be during acceptance testing, installation and periodically 

thereafter.” (see ESD TR20.20 section 5.3.5.3 Testing) 

•  Garments

 - “To maintain process control, it is imperative 

that the garment be tested per ANSI/ESD STM 2.1. The 

point-to-point and sleeve-to-sleeve resistance test should 

be made.”  (see ESD TR20.20 section 5.3.13.3.1.8 

Periodic Testing)

Surface Resistivity

Theoretically, Resistivity is 10 times greater than Resistance.
For example, a material that measures 10

7

 ohms in surface 

resistance should measure 10

8

 ohms/square in surface 

resistivity.*
Ref: ANSI/ESD STM11.11 section 12.0 Conversion to 

Resistivity states, “When it is appropriate to convert a 

resistance obtained by this test method to an equivalent 

resistivity in ohms per square, multiply the resistance 

measurements obtained by this method by 10.  The 

conversion factor of 10 is derived from the geometry of the 

electrode assembly.”
*Note: Although STM11.11 uses ohms/square, per ESD 

TR20.20, the correct unit of measure for resistivity is just 

ohms.

Product Qualification Test Lab Test Procedure Guideline

For test lab use of ESD Worksurfaces, Floor Materials, 

Footwear, Garments or Seating, the best advice is to follow 

the test methods in applicable ESD Association documents 

which include details such as:
•  Cleaning (For example per S4.1, “The test specimens 

and electrodes shall be cleaned twice with a minimum 

70% isopropanol-water solution using a clean, low-linting 

cloth each time).  Allow to the material to dry.

•  Environmental chamber (For example per S4.1, control 

relative humidity to 12 ±3% relative humidity and 50 ±5% 

relative humidity and temperature to 23 ±1°C) 

•  Specimen support surface (For example per S4.1, 

greater than 1 x 10

12

 ohms such as PMMA, PTFE or 

polycarbonate) 

•  Specimen Pre-Conditioning (For example per S4.1, at 23 

±1°C; 3 specimens at 12 ±3% relative humidity for up to 

72 hours minimum, and 3 specimens at 50 ±5% relative 

humidity for up to 72 hours minimum) 

•  Reporting Test Results: For example per S4.1 Reporting 

Test Results, report: 

•  Minimum, median and maximum readings for both

 

resistance-to-groundable point and point-to-point 

 

resistance in ohms at low relative humidity 

•  Minimum, median and maximum readings for both  

 

resistance-to-groundable point and point-to-point  

 

resistance in ohms at moderate relative humidity 

•  Temperature

•  Relative humidity

•  Actual duration of conditioning

•  Test equipment used and latest calibration date

Содержание 19787

Страница 1: ...r ANSI ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person A Compliance Verification Plan shall be established to ensure the Organization s fulfillment of the technical requirements of the ESD Control Program Plan Process monitoring measurements shall be conducted in accordance with a Compliance Verification Plan that identifies the technical requirements t...

Страница 2: ...Test Leads 5 Length 2 Five Pound Electrodes 2 AA Alkaline Batteries 1 Gator Clip 1 Plastic Carrying Case 1 Certificate of Calibration Figure 2 Desco 19787 Digital Surface Resistance Meter Kit Features and Components Figure 3 Desco 19788 Digital Surface Resistance Meter 19788 Digital Surface Resistance Meter 1 Digital Surface Resistance Meter 2 AA Alkaline Batteries 1 Certificate of Calibration FRO...

Страница 3: ... Check Button When set to FULL the meter will cycle through the 15 second period that displays the temperature and relative humidity before the surface resistance mantissa is displayed When set to QUICK the meter will skip the 15 second period and immediately display the measured surface resistance mantissa G Test Button Push and hold this button to make a measurement with the meter Testing in acc...

Страница 4: ...SD cleaner containing no insulative silicone such as Reztore Antistatic Surface Mat Cleaner and re test to determine if the cause of failure is an insulative dirt layer or the ESD worksurface material Figure 5 Rtg test setup Figure 6 Rtt test setup set the electrodes about 10 apart for worksurfaces and about 36 apart for flooring 10 Measure Resistance Point to Point Rtt on the Surface Do not clean...

Страница 5: ...t and sleeve to sleeve resistance test should be made see ESD TR20 20 section 5 3 13 3 1 8 Periodic Testing Surface Resistivity Theoretically Resistivity is 10 times greater than Resistance For example a material that measures 107 ohms in surface resistance should measure 108 ohms square in surface resistivity Ref ANSI ESD STM11 11 section 12 0 Conversion to Resistivity states When it is appropria...

Страница 6: ...ter accurate to 1 F Relative Humidity Meter accurate to 2 Test Leads 99 Isopropyl Alcohol and Cleaning Wipes Setup Test Area needs to be free of any high voltage transformers or power supplies and away from any type of fluorescent lighting or high power lighting Worksurface needs to be covered with a grounded conductive mat at 1 0 x 103 or less Technician needs to be connected to equipment ground ...

Страница 7: ...mperature 75 F 3 F 23 9 C 1 7 C Relative Humidity 10 digits A 1 0 x 1012 20 LED 12 Yellow Mantissa 1 20 0 LED 12 Yellow Mantissa 1 00 20 LED 11 Yellow Mantissa 8 00 B 1 0 x 1011 20 LED 11 Yellow Mantissa 1 20 0 LED 11 Yellow Mantissa 1 00 20 LED 10 Green Mantissa 8 00 C 1 0 x 1010 10 LED 10 Green Mantissa 1 10 0 LED 10 Green Mantissa 1 00 10 LED 9 Green Mantissa 9 00 D 1 0 x 109 10 LED 9 Green Man...

Страница 8: ...of of date of purchase Any unit under warranty should be shipped prepaid to the Desco factory Warranty replacements will take approximately two weeks If your unit is out of warranty call our Customer Service Department at 909 627 8178 Chino CA or 781 821 8370 Canton MA for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to ...

Отзывы: