DESCO WEST
- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178
DESCO EAST
- One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Website:
Desco.com
TB-3062
Page 7 of 8
© 2013 DESCO INDUSTRIES, INC.
Employee Owned
Normalization of the Meter
The temperature inside the testing area needs to be 75°F
±3°F at 40% to 60% relative humidity. The meter needs
to stay at a temperature of 75°F ±3°F for approximately 2
hours for proper readings. The meter cannot be normalized
inside objects, enclosed boxes, containers or cases. The
temperature inside an enclosed case will differ from the
outside temperature. These cases will act as insulators to
the meter. The meter must remain stationary in the testing
area for about 2 hours without any significant changes to the
temperature.
NOTE: Accuracy is measured after normalizing the meter for
a minimum of 2 hours.
Calibration Test Procedure
• Use only the test leads that were supplied with the
meter.
• Clean around the banana jack and stereo jack located at
the top of the meter with 99% isopropyl alcohol. Oil from
human fingers can affect the accuracy of the meter.
• Set the Test Voltage Button to 10V HOLD. Press and
hold the Test Button. The measured voltage between
the two test leads should be 10V ±5%.
• Set the Test Voltage Button to AUTO. Press and hold
the Test Button. The measured voltage between the two
test leads should be 100V ±5%.
• Ensure that the Test Voltage Button is set to AUTO
when performing the following measurements. Always
measure from 1.0 x 10
12
ohms and work your way down
to 1.0 x 10
3
. Never start your measurements from 1.0 x
10
3
.
Temperature:
75°F ±3°F (23.9°C ±1.7°C)
Relative Humidity:
±10 digits
A. 1.0 x 10
12
+20%
LED = 12 Yellow Mantissa 1.20
0%
LED = 12 Yellow Mantissa 1.00
-20% LED = 11 Yellow Mantissa 8.00
B. 1.0 x 10
11
+20%
LED = 11 Yellow Mantissa 1.20
0%
LED = 11 Yellow Mantissa 1.00
-20%
LED = 10 Green
Mantissa 8.00
C. 1.0 x 10
10
+10%
LED = 10 Green
Mantissa 1.10
0%
LED = 10 Green
Mantissa 1.00
-10%
LED = 9 Green
Mantissa 9.00
D. 1.0 x 10
9
+10%
LED = 9 Green
Mantissa 1.10
0%
LED = 9 Green
Mantissa 1.00
-10%
LED = 8 Blue
Mantissa 9.00
E. 1.0 x 10
8
+10%
LED = 8 Blue
Mantissa 1.10
0%
LED = 8 Blue
Mantissa 1.00
-10%
LED = 7 Blue
Mantissa 9.00
F. 1.0 x 10
7
+10%
LED = 7 Blue
Mantissa 1.10
0%
LED = 7 Blue
Mantissa 1.00
-10%
LED = 6 Blue
Mantissa 9.00
G. 1.0 x 10
6
+10%
LED = 6 Blue
Mantissa 1.10
0%
LED = 6 Blue
Mantissa 1.00
-10%
LED = 5 Green
Mantissa 9.00
H. 1.0 x 10
5
+10%
LED = 5 Green
Mantissa 1.10
0%
LED = 5 Green
Mantissa 1.00
-10%
LED = 4 Green
Mantissa 9.00
I. 1.0 x 10
4
+10%
LED = 4 Green
Mantissa 1.10
0%
LED = 4 Green
Mantissa 1.00
-10%
LED = 3 Red
Mantissa 9.00
J. 1.0 x 10
3
+10%
LED = 3 Red
Mantissa 1.10
0%
LED = 3 Red
Mantissa 1.00
-10%
LED = <3 Red
Mantissa <0.90
Specifications
Resistance Ranges
1 x 10
3
- 1 x 10
12
ohms @ 10 Volts,
complies with ANSI/ESD S4.1
1 x 10
6
- 1 x 10
12
ohms @ 100 Volts,
complies with ANSI/ESD S4.1
Accuracy
Resistance measurements within
±10% (±20% for resistance values
1 x 10
11
and greater), complies with
ANSI/ESD-S4.1. Under load
voltages of 10 Volts ± 5% and 100
Volts ± 5% exceeds requirements
of ANSI/ESD S4.1
Power Supply
2 AA alkaline batteries
Battery Life
Approx. 1,500 measurements
Operating Temperature 32°F - 104°F (0 - 40°C)
Meter Dimensions
8.1" x 4.1" x 1.5"
(20.6cm x 10.4cm x 3.8cm)
Meter Weight
0.7 lbs (0.3 kg)
Kit Dimensions
13.5" x 10.8" x 4.0"
(34.3cm x 27.4cm x 10.2cm)
Kit Weight
13.0 lbs (5.9 kg)
External Electrode
5 lbs (±2 oz), 2.5" in diameter,
complies with ANSI/ESD S4.1