TB-3062

 Page 7 of 8

© 2013 DESCO INDUSTRIES, INC.

Employee Owned

Normalization of the Meter

The temperature inside the testing area needs to be 75°F 

±3°F at 40% to 60% relative humidity.  The meter needs 

to stay at a temperature of 75°F ±3°F for approximately 2 

hours for proper readings.  The meter cannot be normalized 

inside objects, enclosed boxes, containers or cases.  The 

temperature inside an enclosed case will differ from the 

outside temperature.  These cases will act as insulators to 

the meter.  The meter must remain stationary in the testing 

area for about 2 hours without any significant changes to the 

temperature.
NOTE: Accuracy is measured after normalizing the meter for 

a minimum of 2 hours.

Calibration Test Procedure

•  Use only the test leads that were supplied with the 

meter. 

•  Clean around the banana jack and stereo jack located at 

the top of the meter with 99% isopropyl alcohol.  Oil from 

human fingers can affect the accuracy of the meter. 

•  Set the Test Voltage Button to 10V HOLD.  Press and 

hold the Test Button.  The measured voltage between 

the two test leads should be 10V ±5%. 

•  Set the Test Voltage Button to AUTO.  Press and hold 

the Test Button.  The measured voltage between the two 

test leads should be 100V ±5%. 

•  Ensure that the Test Voltage Button is set to AUTO 

when performing the following measurements.  Always 

measure from 1.0 x 10

12

 ohms and work your way down 

to 1.0 x 10

3

.  Never start your measurements from 1.0 x 

10

3

.

Temperature: 

 

75°F ±3°F (23.9°C ±1.7°C)

Relative Humidity: 

±10 digits

A. 1.0 x 10

12

 

+20% 

LED = 12 Yellow  Mantissa 1.20 

 

0% 

LED = 12 Yellow  Mantissa 1.00 

 

-20%   LED = 11 Yellow   Mantissa 8.00 

B. 1.0 x 10

11

 

+20% 

LED = 11 Yellow  Mantissa 1.20 

 

0% 

LED = 11 Yellow  Mantissa 1.00 

 

-20% 

LED = 10 Green 

Mantissa 8.00

C. 1.0 x 10

10

 

+10% 

LED = 10 Green 

Mantissa 1.10

 

0% 

LED = 10 Green 

Mantissa 1.00

 

-10% 

LED =   9 Green 

Mantissa 9.00

D. 1.0 x 10

9

 

+10% 

LED =   9 Green 

Mantissa 1.10

 

0% 

LED =   9 Green 

Mantissa 1.00

 

-10% 

LED =   8 Blue 

Mantissa 9.00

E. 1.0 x 10

8

 

+10% 

LED =   8 Blue 

Mantissa 1.10

 

0% 

LED =   8 Blue 

Mantissa 1.00

 

-10% 

LED =   7 Blue 

Mantissa 9.00

F. 1.0 x 10

7

 

+10% 

LED =   7 Blue 

Mantissa 1.10

 

0% 

LED =   7 Blue 

Mantissa 1.00

 

-10% 

LED =   6 Blue 

Mantissa 9.00

G. 1.0 x 10

6

 

+10% 

LED =   6 Blue 

Mantissa 1.10

 

0% 

LED =   6 Blue 

Mantissa 1.00

 

-10% 

LED =   5 Green 

Mantissa 9.00

H. 1.0 x 10

5

 

+10% 

LED =   5 Green 

Mantissa 1.10

 

0% 

LED =   5 Green 

Mantissa 1.00

 

-10% 

LED =   4 Green 

Mantissa 9.00

I. 1.0 x 10

4

 

+10% 

LED =   4 Green 

Mantissa 1.10

 

0% 

LED =   4 Green 

Mantissa 1.00

 

-10% 

LED =   3 Red 

Mantissa 9.00

J. 1.0 x 10

3

 

+10% 

LED =   3 Red 

Mantissa 1.10

 

0% 

LED =   3 Red 

Mantissa 1.00

 

-10% 

LED =   <3 Red 

Mantissa <0.90

Specifications

Resistance Ranges 

1 x 10

3

 - 1 x 10

12

 ohms @ 10 Volts, 

 

complies with ANSI/ESD S4.1

 

1 x 10

6

 - 1 x 10

12

 ohms @ 100 Volts, 

 

complies with ANSI/ESD S4.1

Accuracy 

Resistance measurements within 

 

±10% (±20% for resistance values 

 

1 x 10

11

 and greater), complies with 

 

ANSI/ESD-S4.1.  Under load 

 

voltages of 10 Volts ± 5% and 100 

 

Volts ± 5% exceeds requirements 

 

of ANSI/ESD S4.1

Power Supply 

2 AA alkaline batteries

Battery Life 

Approx. 1,500 measurements

Operating Temperature  32°F - 104°F (0 - 40°C)
Meter Dimensions  

8.1" x 4.1" x 1.5"

 

(20.6cm  x 10.4cm x 3.8cm)

Meter Weight 

0.7 lbs  (0.3 kg)

Kit Dimensions  

13.5" x 10.8" x 4.0"

 

(34.3cm  x 27.4cm x 10.2cm)

Kit Weight 

13.0 lbs  (5.9 kg)

External Electrode 

5 lbs (±2 oz), 2.5" in diameter, 

 

complies with ANSI/ESD S4.1

Содержание 19787

Страница 1: ...r ANSI ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person A Compliance Verification Plan shall be established to ensure the Organization s fulfillment of the technical requirements of the ESD Control Program Plan Process monitoring measurements shall be conducted in accordance with a Compliance Verification Plan that identifies the technical requirements t...

Страница 2: ...Test Leads 5 Length 2 Five Pound Electrodes 2 AA Alkaline Batteries 1 Gator Clip 1 Plastic Carrying Case 1 Certificate of Calibration Figure 2 Desco 19787 Digital Surface Resistance Meter Kit Features and Components Figure 3 Desco 19788 Digital Surface Resistance Meter 19788 Digital Surface Resistance Meter 1 Digital Surface Resistance Meter 2 AA Alkaline Batteries 1 Certificate of Calibration FRO...

Страница 3: ... Check Button When set to FULL the meter will cycle through the 15 second period that displays the temperature and relative humidity before the surface resistance mantissa is displayed When set to QUICK the meter will skip the 15 second period and immediately display the measured surface resistance mantissa G Test Button Push and hold this button to make a measurement with the meter Testing in acc...

Страница 4: ...SD cleaner containing no insulative silicone such as Reztore Antistatic Surface Mat Cleaner and re test to determine if the cause of failure is an insulative dirt layer or the ESD worksurface material Figure 5 Rtg test setup Figure 6 Rtt test setup set the electrodes about 10 apart for worksurfaces and about 36 apart for flooring 10 Measure Resistance Point to Point Rtt on the Surface Do not clean...

Страница 5: ...t and sleeve to sleeve resistance test should be made see ESD TR20 20 section 5 3 13 3 1 8 Periodic Testing Surface Resistivity Theoretically Resistivity is 10 times greater than Resistance For example a material that measures 107 ohms in surface resistance should measure 108 ohms square in surface resistivity Ref ANSI ESD STM11 11 section 12 0 Conversion to Resistivity states When it is appropria...

Страница 6: ...ter accurate to 1 F Relative Humidity Meter accurate to 2 Test Leads 99 Isopropyl Alcohol and Cleaning Wipes Setup Test Area needs to be free of any high voltage transformers or power supplies and away from any type of fluorescent lighting or high power lighting Worksurface needs to be covered with a grounded conductive mat at 1 0 x 103 or less Technician needs to be connected to equipment ground ...

Страница 7: ...mperature 75 F 3 F 23 9 C 1 7 C Relative Humidity 10 digits A 1 0 x 1012 20 LED 12 Yellow Mantissa 1 20 0 LED 12 Yellow Mantissa 1 00 20 LED 11 Yellow Mantissa 8 00 B 1 0 x 1011 20 LED 11 Yellow Mantissa 1 20 0 LED 11 Yellow Mantissa 1 00 20 LED 10 Green Mantissa 8 00 C 1 0 x 1010 10 LED 10 Green Mantissa 1 10 0 LED 10 Green Mantissa 1 00 10 LED 9 Green Mantissa 9 00 D 1 0 x 109 10 LED 9 Green Man...

Страница 8: ...of of date of purchase Any unit under warranty should be shipped prepaid to the Desco factory Warranty replacements will take approximately two weeks If your unit is out of warranty call our Customer Service Department at 909 627 8178 Chino CA or 781 821 8370 Canton MA for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to ...

Отзывы: