background image

TB-3040

 Page 4 of 6

© 2009 DESCO INDUSTRIES, INC.

Employee Owned

Specifications

Measurement Range (switch selectable)

Low Range: 

 

0 to ±1.99 kV / inch

High Range: 

 

0 to ±19.99 kV / inch

Measurement Accuracy

Voltage Monitor Output: > ±5% of reading ±10 mV

Voltage Display:  

> ±5% of reading ±2 counts

Measurement Stability

±10 counts

Automatic Shutoff

Unit will shut off after 20 minutes after last switch activity

Power Requirements

One (1) 9V alkaline battery

Operating Time

Greater than 50 hours, with new battery at 21°C 

continuous usage

Operating Conditions

Temperature:   

10°C to 30°C

Relative Humidity: 

Up to 80%, non-condensing

Altitude: 

 

Up to 2,000 meters

Dimensions

0.94" H x 2.75" W x 4.94" L

(23.9mm H x 69.9mm W x 125.5mm L)

Weight (with battery)

4.9 oz

(140 g)

Voltage Monitor Connection

2.5mm audio jack

CE Certified

Ionization Test Kit Accessories

Desco offers accessories for the Digital Static Field 

Meter designed to facilitate routine auditing and periodic 

testing of ionization equipment (Ref: ANSI/ESD SP3.3). 

The Meter and accessories combination can be used to 

test an ionizer’s overall performance.  This highly 

portable test kit allows the user to make quick and 

accurate offset voltage balance level and neutralization 

discharge time measurements (counting or using 

stopwatch).  These accessories are available as Desco 

item# 

19496

 Test Kit Upgrade which includes the 

Conductive Plate and Charger.

Figure 5. Installing the 19498 Conductive Plate

12"

18"

Figure 6. Auditing ionization equipment with the Digital 
Static Field Meter and Conductive Plate (Ref: ANSI/ ESD 
SP3.3)

Desco 

19448

 Personal HBM Test Fixture

Note:

 Desco 

19498

 Conductive Plate required

The Desco 

19448

 Personal HBM Test Fixture turns your

model 

19492

 Digital Field Meter into a Human Body 

Model tribo-electric charging tester.  This kit will allow 

you to measure electrical charges generated on the 

human body while walking or moving across an 

insulative or conductive surface.

Содержание 19492

Страница 1: ...If an ESDS ESD sensitive device is placed in that electrostatic field a voltage may be induced on the device If the device is then momentarily grounded a transfer of charge from the device occurs as a CDM Charged Device Model event If the device is removed from the region of the electrostatic field and grounded again a second CDM event will occur as charge of opposite polarity from the first event...

Страница 2: ...e meter to the 2 kV 3 decimal places range Point the top of the Meter approximately 1 inch away from a grounded metal surface Use the red LED range guide The Meter is properly positioned when the projected red bullseyes are centered on top of each other Press and hold the RANGE ZERO button until the meter displays 000 MAKING A MEASUREMENT Place the meter 1 inch from the object to be measured This ...

Страница 3: ...s advised to do this between measurements to prolong battery life ANALOG OUTPUT The analog output jack labeled OUT on the face of the meter accepts a standard 2 5 mm monaural phone plug and is provided so the output of the Digital Static Field Meter may be connected to an oscilloscope strip chart recorder external meter or other device Use the included cord to achieve a connection between the fiel...

Страница 4: ...gital Static Field Meter designed to facilitate routine auditing and periodic testing of ionization equipment Ref ANSI ESD SP3 3 The Meter and accessories combination can be used to test an ionizer s overall performance This highly portable test kit allows the user to make quick and accurate offset voltage balance level and neutralization discharge time measurements counting or using stopwatch The...

Страница 5: ...the Installation and Removal section Once installed on the Field Meter the HBM Test Fixture itself is isolated from the meter case and ground and should remain that way Cradle the Test Fixture in the palm of your hand Once the Meter is grounded and you are standing on the substrate to be tested simply walk around shuffle your feet raise a foot or use the walking pattern per Figure 1 of ANSI ESD ST...

Страница 6: ...approximately two weeks If your unit is out of warranty call our Customer Service Department at 909 627 8178 Chino CA or 781 821 8370 Canton MA for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUC...

Отзывы: