TB-3040

 Page 2 of 6

© 2009 DESCO INDUSTRIES, INC.

Employee Owned

Packaging

1  Digital Static Field Meter

1  9V Alkaline Battery

1  Data Output Cord

1  Certificate of Calibration

Features and Components

C. Analog Output Jack:

 A low-voltage replica of the 

measured voltage is provided at this output.  The voltage 

at this output is 1/1000th (±2 kV range) or 1/10,000 (±20 

kV range) of the measured voltage.

D. POWER Button:

 Press to turn the unit ON and OFF.

E. 4mm Stud:

 Use this stud to ground the Meter using 

a 4mm ground cord (not provided - any standard single 

wire wrist strap cord can be used).

F.  Battery Cover:

 Slide the cover down to open the 9V 

battery compartment.

Operation

Note: 

The 

19492

 Digital Static Field Meter is built in a 

conductive case.  The instrument senses the difference 

in potential between the case (and the person holding 

the case / ground connection) and the surface under 

test.  Ensure that the person using the instrument is 

grounded or that the rear panel ground snap connection 

is utilized to achieve accurate measurements.

BATTERY CHECK

The battery should be replaced when “BAT” is indicated 

on the display.  Always replace the battery with a 9V 

alkaline or equivalent battery in order to remain CE 

compliant.

ZERO THE METER

Turn the meter on by pressing the POWER button.  

Press the RANGE / ZERO button to set the meter to 

the 2 kV (3 decimal places) range.  Point the top of the 

Meter approximately 1 inch away from a grounded metal 

surface.  Use the red LED range guide.  The Meter is 

properly positioned when the projected red bullseyes 

are centered on top of each other.  Press and hold the 

RANGE / ZERO button until the meter displays “.000”.

MAKING A MEASUREMENT

Place the meter 1 inch from the object to be measured.  

This distance is measured from the front edge of the 

meter case to the surface of the object.  The meter now 

displays a reading (from 0 to ±.200 or ±2.00) of the 

electrostatic field in kilovolts per inch.

Note:

 The red ranging lights are provided to help place 

the meter at the correct distance from the object.  The 

lights are set to produce a concentric red bullseye 

pattern on a flat opaque surface 1 inch from the front 

edge of the meter.  This can be practiced by aiming the 

meter at a sheet of white paper.
The display will indicate “1” or “-1” when the meter is 

over-ranged.  Change the range of the unit if necessary.  

If the measurement exceeds 20 kV, move the meter 

farther away from the object and multiply the reading 

by the distance (in inches) away from the object being 

measured.  The measurement accuracy is dependent on 

Figure 2. Digital Static Field Meter features and 
components

A

B

C

D

FRONT VIEW

BACK VIEW

E

F

A. HOLD Button:

 Press to freeze the reading on the 

display.  Press again to return to normal measurement 

operation.

B. RANGE / ZERO Button:

 Press to select the 

measurement range.  Press and hold to zero the meter.

Содержание 19492

Страница 1: ...If an ESDS ESD sensitive device is placed in that electrostatic field a voltage may be induced on the device If the device is then momentarily grounded a transfer of charge from the device occurs as a CDM Charged Device Model event If the device is removed from the region of the electrostatic field and grounded again a second CDM event will occur as charge of opposite polarity from the first event...

Страница 2: ...e meter to the 2 kV 3 decimal places range Point the top of the Meter approximately 1 inch away from a grounded metal surface Use the red LED range guide The Meter is properly positioned when the projected red bullseyes are centered on top of each other Press and hold the RANGE ZERO button until the meter displays 000 MAKING A MEASUREMENT Place the meter 1 inch from the object to be measured This ...

Страница 3: ...s advised to do this between measurements to prolong battery life ANALOG OUTPUT The analog output jack labeled OUT on the face of the meter accepts a standard 2 5 mm monaural phone plug and is provided so the output of the Digital Static Field Meter may be connected to an oscilloscope strip chart recorder external meter or other device Use the included cord to achieve a connection between the fiel...

Страница 4: ...gital Static Field Meter designed to facilitate routine auditing and periodic testing of ionization equipment Ref ANSI ESD SP3 3 The Meter and accessories combination can be used to test an ionizer s overall performance This highly portable test kit allows the user to make quick and accurate offset voltage balance level and neutralization discharge time measurements counting or using stopwatch The...

Страница 5: ...the Installation and Removal section Once installed on the Field Meter the HBM Test Fixture itself is isolated from the meter case and ground and should remain that way Cradle the Test Fixture in the palm of your hand Once the Meter is grounded and you are standing on the substrate to be tested simply walk around shuffle your feet raise a foot or use the walking pattern per Figure 1 of ANSI ESD ST...

Страница 6: ...approximately two weeks If your unit is out of warranty call our Customer Service Department at 909 627 8178 Chino CA or 781 821 8370 Canton MA for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUC...

Отзывы: