Remove the MegaPeel from the room when performing steam treatments. Steam remains in the air even
after the steam treatment is complete
IF STEAM STILL SEEMS TO BE AN ISSUE:
After your last treatment of the day:
Take the short blue tube from the spare parts kit.
Connect it to the blue crystal tube and to the pink tube on the waste filter
Leave the blue tube connected overnight.
Run the system for several minutes in the morning before replacing the crystal canister. This will
introduce dry air into the system.
If after doing all the above clearings and crystals are not coming out of
the hand-piece, call DermaMed’s Technical Support line, 610.358-4447.
We will help you make the machine work properly.