background image

Gold

  

Series

TROUBLESHOOTING

We expect that your experiences with the MegaPeel product and process will always run smoothly. In 
case you run into problems, here’s a handful of suggestions.

WHAT IF THE MACHINE DOES NOT TURN ON?

Make sure that cord in the back is plugged in and the power button on the front of the machine is on.
Check the fuse.
Unplug the machine.
On the back of the machine, right of the circular vent, is the black fuse box.
Use a regular, flat-blade screwdriver to pull out the front of the black fuse box.
Again using the screw driver, pop off the red fuse box. If the fuse looks blown, replace it with a new fuse.
From the spare parts kit, take a new fuse, which is a glass-and-steel tube about 1 1/4-inch long.
Using your fingers, remove the old fuse from the MegaPeel. 
Install the new fuse. 
Throw the old fuse into regular trash.

WHAT IF CRYSTALS ARE NOT COMING OUT OF THE HAND-PIECE? #1

Remove the lid.
Lift the top of the machine.
Lift the white crystal container (the smaller of the 2 containers) from the machine.
To remove the blue tube, push down with your thumb on the small, white plastic piece located on the 
crystal container. 

Before troubleshooting your MegaPeel be sure that 

the crystal cannister is                                                    

not empty or the waste filter is not full.

Содержание Gold Series

Страница 1: ...sing the screw driver pop off the red fuse box If the fuse looks blown replace it with a new fuse From the spare parts kit take a new fuse which is a glass and steel tube about 1 1 4 inch long Using your fingers remove the old fuse from the MegaPeel Install the new fuse Throw the old fuse into regular trash WHAT IF CRYSTALS ARE NOT COMING OUT OF THE HAND PIECE 1 Remove the lid Lift the top of the ...

Страница 2: ... the front of the machine Increase the vacuum pressure by turning the knob all the way to the right Turn on the machine Let the machine run for one minute When the machine stops remove the short blue tube that you just installed Return the short blue tube to the spare parts kit Reconnect the long blue tube to the crystal container Reconnect the pink tube to the pink connector inside the machine Ru...

Страница 3: ...T IF CRYSTALS ARE STILL NOT COMING OUT OF THE HAND PIECE 3 Remove the lid Check the pink and blue tubing which you attach the hand piece to and see if there is a clog If you see a clog replace the tubing If you do not see a clog disconnect the hand piece Use the needle in the spare parts kit to unclog the holes Knock the hand piece forcefully against a hard surface If you still do not see crystals...

Страница 4: ...tops remove the short blue tube that you just installed Return the short blue tube to the spare parts kit Reconnect the long blue tube to the crystal container Reconnect the pink tube to the waste filter connection Run the machine The MegaPeel should now be receiving crystals IF YOU HAVE A STEAMER IN THE ROOM Steam can cause clogging problems We do not recommend keeping your machine in the same ro...

Страница 5: ... spare parts kit Connect it to the blue crystal tube and to the pink tube on the waste filter Leave the blue tube connected overnight Run the system for several minutes in the morning before replacing the crystal canister This will introduce dry air into the system If after doing all the above clearings and crystals are not coming out of the hand piece call DermaMed s Technical Support line 610 35...

Отзывы: