DERMALOG FLC1 Light Edition User Guide
17
/
24
To achieve optimal results, follow the recommendations from DERMALOG:
Do NOT point the DERMALOG FLC1 Light Edition at a heat source because this may lead to
inaccurate measurements of the person’s body temperature. The person must also NOT directly
be exposed to radiation from heat sources (e.g. from a heater in the room).
Make sure that light sources (e.g. windows) are NOT behind the person. Position the camera
such that light sources are behind the camera to ensure frontal lighting.
The DERMALOG FLC1 Light Edition is positioned at the person’s eye level.
The distance between the person and the DERMALOG FLC1 Light Edition is between 0.5 m and 2 m.
Figure 6: Recommended alignment of DERMALOG FLC1 Light Edition
7.2
Connect the hardware
Prerequisites:
You comply with the safety instructions (see chapter 3).
You have positioned the DERMALOG FLC1 Light Edition according to chapter 7.1
If you want to connect the DERMALOG FLC1 Light Edition via a USB hub, always use a self-
powered (active) hub.
START:
STEP 1
Check the DERMALOG FLC1 Light Edition and the USB cables for damages and breaks.
If there are any damages and breaks, stop using the DERMALOG FLC1 Light Edition and contact
DERMALOG via email to [email protected].
Otherwise, proceed with Step 2.
STEP 2
Plug the USB-C connector of the USB 3.0 cable into the back port of the DERMALOG FLC1 Light
Edition.
The USB-C-connector is the smaller connector on the USB 3.0 cable.