4
UNIT SPECIFICATIONS
DIMENSIONS
Height: .................... 12-1/8 in. (30.8 cm)
Width: .................... 7 in. (17.8 cm)
Depth: .................... 10-3/4 in. (27.3 cm)
Power Cord:............ 62 in. (157.5 cm)
WEIGHT
Domestic: .............. 10 lbs. (4.5Kg)
Export: .................. 15.5 lbs. (7.03 Kg)
VOLTAGE/CURRENT .. 100V/2.7 AMP
115V/2.35 AMP
220-240V/1.2 AMP
FREQUENCY ................ 50/60 Hz
LAMP .......................... Special Tungsten Halogen
(See Parts Listing section)
TRIAD
®
2000 LIGHT CURING UNIT
INSTALLATION
UNPACKING AND INSTALLATION
DO NOT CONNECT POWER CORD TO
OUTLET AT THIS TIME.
1. Remove unit from shipping carton and place
in the desired operating location.
2. Open the processing chamber door and
remove packing material.
NOTE
: Save shipping and packing mate-
rials. If service is required at any time on
your Triad 2000 unit, you are respon-
sible for its safe* shipment back to
DENTSPLY. This is best accomplished
if the original shipping carton and pack-
ing materials are used.
3. Connect the POWER cord to an appropriate
power source.
4. Turn the POWER switch ON. The cooling
fan should operate.
5. With the access door closed, set the TIMER
dial to 1 minute. Depress the CURE switch
to ON. The lamp should illuminate.
6. Depress the TABLE switch to ON. The table
should rotate.
1. Lamp Access Door Thumbscrew
2. Lamp Mounting
3. Lamp Access Door
4. Optimum Curing “Sightline”
*Supply OSHA information re: Bloodborne pathogens policy on
shipping possible contaminated equipment for service.
FIGURE 2