21
Guidance And Manufacturer’s Declaration - Electromagnetic Emissions
The Model G136 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Model G136 should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance
Level
Electromagnetic environment - guidance
Radiated RF
IEC 61000-4-3
3 V/m
26 MHz to 2.5 GHz
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the Model G136,
including cables, than the Recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
d = 1.7 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the -
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
a
should be less than
the compliance level in each frequency range
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Model G136 is used
exceeds the applicable RF compliance level above, the Model G136 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the Model G136.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Содержание Cavitron Plus Ultrasonic Scaler
Страница 1: ......
Страница 3: ......
Страница 4: ......
Страница 11: ......
Страница 12: ......
Страница 13: ......
Страница 14: ......
Страница 15: ......
Страница 19: ......
Страница 25: ......
Страница 26: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 37: ......
Страница 38: ......
Страница 43: ......
Страница 49: ......
Страница 50: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 66: ......
Страница 73: ......
Страница 74: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......
Страница 90: ......
Страница 96: ......
Страница 97: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......
Страница 108: ......
Страница 109: ......
Страница 113: ......
Страница 120: ......
Страница 121: ......
Страница 129: ......
Страница 130: ......
Страница 131: ......
Страница 132: ......
Страница 133: ......
Страница 138: ......
Страница 143: ...143...
Страница 144: ......