15
Electric Heat Pluggers & Thermal Response Tip
Electric Heat Pluggers (EHP) and Thermal Response Tip (TRT) must be
disinfected and sterilized before every use:
1.
Gently use a soft brush for removal of debris.
2.
Clean the EHPs and TRTs under running water for a minimum of 1
minute using water and a mild, non-abrasive, detergent or
disinfectant.
3.
Do
NOT
use disinfecting solutions which contain phenol, anti-
corrosives, very acidic or very alkaline solutions.
4.
Rinse the EHPs and/or the TRT under running water for a minimum of
1 minute to remove all chemicals.
5.
Dry thoroughly.
6.
Steam autoclave Electric Heat Pluggers and Thermal Response Tips for
15 minutes at 132°C before initial use and between each patient.
7.
Inspect the EHPs and TRTs regularly for damage.
•
Deformed or oxidized pluggers and tips should be replaced.
•
All EHPs and TRTs will gradually lose their heating efficiency over a
period of time.
•
Use of damaged or contaminated tips is at the user’s own risk. All
liability is excluded in this case.
8.
Dispose of EHPs and TRTs in a sharps receptacle.
NOTE:
Observe the concentration of the cleaning or disinfecting
solution stated by the manufacturer!
WARNING:
Do not touch Electric Heat Pluggers while hot!