E. Connect wire # 7 as follows:
F. Filament Voltage Adjustments Replace the front panel.
a. Momentarily operate the generator and observe Filament Voltage Meter.
b. Filament voltage should read between 12.3 and 12.9 VAC (Fig 9). If lower than this range,
place wire 6 at the transformer, to “L”. If higher than this range, place wire 6, to “H”.
G. Grid Coil Adjustments (Fig 9)
a. Place 2 ingots of Vitallium in a new crucible.
b. Switch the power level switch (Fig 12-3) to position 3.
c. Operate generator for 3 seconds and record on a piece of paper the grid current
Electromatic - 4
9
VOLTAGE AT
FUSE BOX
CONNECT
WIRE # 7 TO:
195-205 V
200
206-215 V
210
216-225 V
220
226-235 V
230
236-245 V
240
246-255 V
250
FIG. 8
FIG. 9
WARNING:
Before any grid coil adjustments are made, press the Stop button to
avoid melting the two test ingots.