116
Guidance and manufacturer´s declaration – Electromagnetic Immunity
Table 5, Part A according to EN 60601-1-2: 2015 (ed.4)
The SmartLite
®
Focus
®
is intended for use in the electromagnetic environment specified below.
The customer or the user of SmartLite
®
Focus
®
should assure that it is used in such an environment.
Phenomenon
EN 60601-1-2
Immunity test level
Compliance level
Electromagnetic
environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be
wood, concrete or ce-
ramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 20%.
Electrical fast tran-
sient disturbances/
bursts
IEC 61000-4-4
± 2 kV (100 kHz)
for power supply lines
± 1 kV (100 kHz)
for input/output lines
± 2 kV (100 kHz)
for power supply lines
± 1 kV (100 kHz)
for input/output lines
Surge
IEC 61000-4-5
± 0,5 kV, ± 1 kV line(s)
to line(s)
± 0,5 kV, ± 1 kV, ± 2 kV
line(s) to earth
± 0,5 kV, ± 1 kV line(s)
to line(s)
± 0,5 kV, ± 1 kV, ± 2 kV
line(s) to earth
Voltage dips an
variations on power
supply input lines
IEC 61000-4-11
0% U
T
for 0,5 cycle
(at 0, 45, 90, 135, 180,
225, 270, 315 degrees)
0% U
T
for 1 cycle
and
70% U
T
for 25 cycles (50 Hz)
or 30 cycles (60 Hz)
single-phase: 0 degree
0% U
T
for 0,5 cycle
(at 0, 45, 90, 135, 180,
225, 270, 315 degrees)
0% U
T
for 1 cycle
and
70% U
T
for 25 cycles (50 Hz)
or 30 cycles (60 Hz)
single-phase: 0 degree
Short voltage inter-
ruptions on power
supply input lines
IEC 61000-4-11
0% U
T
for 250 cycles (50 Hz)
or 300 cycles (60 Hz)
0% U
T
for 250 cycles (50 Hz)
or 300 cycles (60 Hz)
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.