6
3 Step-by-step instructions
3.1 Installation and charging (
Procedure A
and
Procedure B
)
1. Remove floor panel
3.1
of charging base by lifting floor panel at its back end (see
Figure B2
)
.
2. Connect jack
4.1
to charging base socket
3.2
(see
Figure B3
).
3. Close floor panel (see
Figure B4
) and plug in power connector
4
.
4. Make sure power socket used for power connector is accessible at all times in case
of necessary emergency disconnection.
5. Insert battery pack
2
into SmartLite
®
Focus
®
handpiece
1
(see
Figure A2
and
Figure A3
).
SmartLite
®
Focus
®
battery pack is pre-charged and ready to use upon purchase.
6. To recharge the battery, insert the handpiece
1
into the charging base
3
(see
Figure B5
). The battery light will be flashing green, indicating battery is re-
charging.
The SmartLite
®
Focus
®
Automatic Quick Charging Mode allows a minimum of 5 minutes
operating time after 10 minutes recharging.
3.2 Operation
1. Protect LED tip from gross contamination by applying the single use SmartLite
®
Sleeves (see
Procedure C
).
2. Use the SmartLite
®
Focus
®
Shields supplied with the unit in combination with mounted
SmartLite
®
Sleeves or use suitable, light filtering safety glasses (see
Procedure D
).
Always make sure SmartLite
®
Focus
®
Shields are securely fixed to SmartLite
®
Focus
®
to avoid accidental aspiration (press SmartLite
®
Focus
®
Shield firmly into cor-
rect position). Always make sure SmartLite
®
Focus
®
Shields are properly mounted
on SmartLite
®
Focus
®
without covering the light aperture (see
Figure D3
and
Figure D4
).
3. Adjust the LED tip: The LED tip is rotatable by 330°, thus the position of light emission
may be individually adjusted. The LED tip must be adjusted to an angle that allows
optimal curing of the restoration. The LED tip should be positioned
as closely as
possible to the restoration
.
4. Polymerization: Briefly press the ON/OFF key [1.3.1 ON/OFF key]. Audible operation
signals are described in [1.3.3 Audible signals].
5. To stop curing light before the end of the 20 second cycle, press ON/OFF key at any
time. To cure a restorative material that requires a curing time longer than 20 sec-
onds, repeat activation by pressing the ON/OFF key after the end of each 20 second
curing cycle (see also [2.2 Precautions]).
6. Curing times for a range of Dentsply Sirona materials are outlined in the Curing
Guide provided. When using other products, please refer to respective product’s
Directions for Use.