5 Operation
Sirona Dental Systems GmbH
5.1 Replacing the instrument
Operating Instructions Midwest E Plus
5
Operation
Use only undamaged burs
NOTICE! Use only burs and diamond polishers that are sharp and
undamaged. Use clean burs and diamond polishers to avoid dirt in the
clamping system.
Clamping the bur _ instrument
CAUTION! A loose or partially removed bur instrument can come loose
from the head or break off. Risk of injury! Therefore the instrument must
only be used when the bur instrument is securely clamped at least 10 mm
deep.
Only operate when the clamping system is closed
CAUTION! The handpiece must only be operated with the clamping
system closed.
Overheating at the preparation site
CAUTION! Insufficient cooling leads to overheating of the preparation
site and damage to the tooth substance. Ensure that the water content is
> 50ml/min.
Zurückziehen der Wange _ Instrument
CAUTION! Do not pull the patient's cheek back with the contra-angle
handpiece while the motor is running. This would actuate the pushbutton,
thus creating a risk of burning the patient's oral mucosa.
Red contra-angle handpiece
CAUTION! Always operate the contra-angle handpieces with cooling
spray when marked red!
Observing manufacturer documentation
IMPORTANT: For application details and operating data of the different
preparation tools, please refer to the information provided by the
manufacturer.
5.1
Replacing the instrument
Motor at standstill
CAUTION! The instrument should only be fitted or removed when the
motor is at standstill.
Instruments with light only on motors with light
NOTICE! Do not use instruments with light on motors without light.
Attaching the Midwest E Plus instrument
with ISO slider
✔
The motor is at a standstill.
1. Insert the instrument until it snaps into place.
2. When using a motor with a light groove (A): turn the instrument until
the ISO slider clicks into place.
Removing the instrument
✔
The motor is at a standstill.
➢ Detach the instrument. Do not pull on the supply hose while doing
this.
5.2
Inserting and removing bur instruments
Free movement of the pushbutton
IMPORTANT: Check the push button to make sure it moves freely!
Use only undamaged burs
NOTICE! Use only burs and diamond polishers that are sharp and
undamaged. Use clean burs and diamond polishers to avoid dirt in the
clamping system.
Clamping the bur _ instrument
CAUTION! A loose or partially removed bur instrument can come loose
from the head or break off. Risk of injury! Therefore the instrument must
only be used when the bur instrument is securely clamped at least 10 mm
deep.
Содержание Midwest E Plus
Страница 2: ......
Страница 4: ......
Страница 5: ...k W l f MUKOMNS j b m b b Operating Instructions...
Страница 27: ...k W k D MUKOMNS j b m c c Notice d utilisation...
Страница 49: ...k W j MUKOMNS j b m b b Manual del operador...
Страница 71: ......