7
Symptom/Cause
Corrective Measures
Intermittent activation or no
activation when stepping on
the Foot Control:
1. Foot Control batteries are
weak.
2. Foot Control too far from
Cavitron
®
300 Series scaler
unit.
3. Foot control is not
synchronized to the scaler.
1. a. Check Foot Control battery condition as indicated by the battery display.
b. Connect the Auxiliary Foot Control cable between the Foot Control and unit.
The unit can be operated with the cable until the batteries are fully charged.
Full charge will take two hours.
2. The distance between the Foot Control and Scaler should be within 10 feet.
Relocate Cavitron
®
300 Series scaler unit.
3. a. Follow the Cavitron
®
300 Series scaler Directions for Use and Service Manual
instructions for Foot Control synchronization.
b. Connect the Auxiliary Foot Control cable between the Foot Control and the
Scaler. The Scaler can be operated with the cable until the Foot Control is
synchronized.
c. Replace the Foot Control. Follow the Cavitron
®
300 Series scaler Directions for
Use and Service Manual instructions for Foot Control Synchronization.
Boost Power mode does not
activate. Diagnostic Display
does not indicate “Boost”
mode:
1. Foot Control not fully
depressed.
2. The Foot Control is
defective.
1. Depress the Foot Control fully. The “Boost” indication should be displayed.
2. Replace the Foot Control. Follow the Cavitron
®
300 Series Directions for Use and
Service Manual instructions for Foot Control synchronization.
Inserts cannot be installed in
the handpiece properly:
1. O-ring on the insert is dry.
2. Incorrect or damaged O-ring
installed on the insert.
3. Metal P-Style inserts cannot
be used with Steri-Mate™
(360) handpieces.
1. Lubricate the O-ring with water. If the O-ring is swollen, worn or frayed replace it.
2. Replace the insert O-ring with Cavitron
®
O-rings. O-rings are available in packs
of 12. Green O-Rings P/N 62605, Black O-Rings P/N 62351.
3. Use Cavitron inserts with plastic or soft grips or use regular Steri-Mate™
handpieces.
Inserts slide out of Steri-
Mate™ (360) when scaling:
1. Water Regulator / Solenoid
set too high.
2. Handpiece water flow set
too high.
3. Insert O-ring worn.
1. Adjust Water Regulator / Solenoid to specifications.
2. Set Handpiece water flow control to approximate 35 cc/min.
3. Replace O-rings.
1. G310 Cavitron
®
300 Series Scaler Troubleshooting And Analysis