Symptom/Cause

Corrective Measures

Handpiece heats up:
 

1.  Insufficient water to cool 

handpiece.

2. Air trapped in the handpiece.

3.  Water filter is clogged with 

debris.

4.  Insert water passageway 

clogged.

5.  Handpiece cable not 

supported during procedure.

6. Worn insert being used.

7.  Operatory water supply 

temperature is high.

1. Increase the setting on the handpiece lavage control until handpiece runs cool.

2.  When the inserts are installed in the handpiece, hold the handpiece in an upright 

position until the trapped air is removed and the water flows properly.

3. Replace the water filter.

4. Replace the Cavitron

®

 insert and check operation.

5.  Loop handpiece cable around arm or support with finger to prevent water 

restriction.

6.  Replace with a new Cavitron

®

 insert.  Worn inserts require higher power settings 

producing more heat.

7. Flush water system before use or use a DualSelect Dispenser as a water source. 

Insert vibrates but no water or 
insufficient water flows from 
the handpiece:
 

1.  Low incoming dental office 

water pressure.

2.  Water filter clogged.   

(Is Water Filter Icon lit?)

3.  Handpiece cable water 

tubing and wires twisted.

4.  Damaged handpiece cable 

flow control.

5.  Obstruction or mineral 

deposits in the water system 
in the unit.

6.  Water Regulator / Solenoid 

out of adjustment.

1.  Measure water pressure at dental office. Adjust incoming source water pressure to 

specification. Water pressure should be 20-40 psi.

2. Replace the water filter when discolored or restriction occurs.

3. Remove restriction if possible or replace handpiece cable assembly.

4. Replace handpiece cable assembly.

5.  a.  Remove the insert and turn the handpiece flow control fully open. Observe the 

water flow. If the flow is good then the obstruction is in the insert.

    b.  If the obstruction is not in the insert, then remove the handpiece water line at 

solenoid and check the water flow. If flow is good, then the obstruction is in 
handpiece tubing.

6.  Adjust water regulator pressure according to Dentsply Sirona Preventive SBU – 

Product Service SOP PS-00211.

No water flow from handpiece 
without insert installed:
 

1.  High dental office water 

pressure.

1.  Install a water pressure regulator on the main water supply line and reduce the 

pressure to 20-40 psi.

1. G310 Cavitron

®

 300 Series Scaler Troubleshooting And Analysis

Содержание Cavitron 300 Series

Страница 1: ...Service Manual Please read carefully and completely before operating unit Cavitron 300 Series Ultrasonic Scaling System...

Страница 2: ...4 G Drive PCB Stack PSOC PC Board Upper PCB 14 15 H Drive PCB Stack Power PC Board Lower PCB 15 16 I Water Regulator Solenoid Replacement 16 17 J USB Connector Assembly Replacement 17 K Handpiece Rece...

Страница 3: ...alled in a confined area such as a cabinet or is too close to a heat source to insure proper air circulation around unit 2 Power Supply malfunction 1 Power Off scaler to allow the internal temperature...

Страница 4: ...nect the ohmmeter between RED GRN wire terminals Flex the handpiece cable and check for intermittent readings If the ohmmeter reading is not consistent or it is indicating an open circuit the handpiec...

Страница 5: ...er clogged Is Water Filter Icon lit 3 Handpiece cable water tubing and wires twisted 4 Damaged handpiece cable flow control 5 Obstruction or mineral deposits in the water system in the unit 6 Water Re...

Страница 6: ...e handpiece while in operation 1 O ring worn on insert 2 Water leak in handpiece water tubing at handpiece or inside the Steri Mate 360 Handpiece 1 Replace the O ring with genuine Cavitron O rings O r...

Страница 7: ...ot Control Synchronization Boost Power mode does not activate Diagnostic Display does not indicate Boost mode 1 Foot Control not fully depressed 2 The Foot Control is defective 1 Depress the Foot Cont...

Страница 8: ...able assembly 4 Return Cavitron 300 Series scaler to Dentsply Sirona Preventive SBU Product Service for factory certified service Service Indicator 2 is displayed in the upper left corner of the main...

Страница 9: ...and hold the selected memory icon 3 Wait for the audible signal The power level should be indicated in the memory location icon Purge mode cannot be activated 1 Purge Icon not depressed properly 2 Foo...

Страница 10: ...ontrol b If the Foot Control power is OFF pressing the Sync button or trying to Sync with the Scaler will not turn the Foot Control ON Lock Icon cannot be activated 1 Icon not depressed correctly 1 Lo...

Страница 11: ...ut the water hose 4 If the Auxiliary Foot Control cable is being used unplug it from the front of the unit 5 Place the unit on a clean non abrasive surface to prevent damage to the cabinet 6 Use a scr...

Страница 12: ...e wires with the Toroid should be connected on the right side The terminals must be fully seated 7 On the back of the cover assembly plug the two wire connector into J1 and the 5 wire connector into J...

Страница 13: ...b fitting on the Burkhart water regulator solenoid Be sure tubing is fully seated 22 Plug the black and white wire terminals with the toroid coil to the Power Entry Module The Black wire goes on the l...

Страница 14: ...screws Torque them to 2 0 5 in lbs 7 Install the Black White wire connector and the Red Black wire connector to the power supply assembly Ensure the connectors are fully seated 8 On the back of the co...

Страница 15: ...for Top Cover Removal 2 Ensure that the power cord has been unplugged from the rear power entry module 3 Unplug the 5 wire connector front right from J13 on the PSOC PC Board 4 Carefully cut the nylo...

Страница 16: ...the Mounting Plate Sub assembly must be slightly raised Remove the three 3 6 x 1 2 Hi Lo Torx screws from the Mounting Plate Sub assembly Also remove the ground wire screw at the top of the metal brac...

Страница 17: ...assembly plug the two wire connector into J1 and the 5 wire connector into J2 from the FPGA PC board Ensure the two wire harnesses are tied together with a nylon cable tie 6 Plug the power cord into t...

Страница 18: ...harnesses are tied together with a nylon cable tie 15 Plug the power cord into the power entry module 16 Reconnect the blue water hose to the push lock water fitting 17 Power up the unit using the pow...

Страница 19: ...sses are tied together with a nylon cable tie 15 Align and place cover assembly on the scaler unit 16 Plug the power cord into the power entry module 17 Reconnect the blue water hose to the push lock...

Страница 20: ...e connector and press the black tab to lock it in place 15 Carefully and fully insert the orange ribbon cable in the connector J3 and press the black tab to lock it in place 16 On the back of the cove...

Страница 21: ...21 NOTES...

Страница 22: ...sembly 25 Base Assembly 26 27 Display FPGA PC Board Assembly 28 Power Supply Module 29 Water Regulator Solenoid 30 Master Ultrasonic PC Board Wire Harness 31 Inlet Harness with Toroid and Wire Harness...

Страница 23: ...SCALER GENERATOR ASSEMBLY ITEM QUANTITY PART NO DESCRIPTION 1 1 Base Assembly G310 2 1 Bezel Sub Assembly G310 3 1 81749 Water and Power Label G310 4 1 81834 1 Steri Mate 360 Handpiece 5 1 82507 1 Ha...

Страница 24: ...310 CAVITRON 300 SERIES SCALER BEZEL SUB ASSEMBLY ITEM QUANTITY PART NO DESCRIPTION 1 1 81746 Bezel Housing G310 2 1 81747 2 Handpiece Holder G310 3 1 81748 Rear Housing G310 4 1 82517 Display Sub Ass...

Страница 25: ...TY PART NO DESCRIPTION 1 1 81741 Mounting Plate G310 2 1 629205005 Water Regulator Solenoid Burkert 3 1 82529 1 Power Supply Assembly Protek Power 4 1 82530 Drive PC Board Stack 5 2 586172005 6 x 1 2...

Страница 26: ...26 3 Cavitron 300 Series Scaler Service Parts Section...

Страница 27: ...ubber Feet Self Adhesive 8 4 586172005 6 x 1 2 Hi Lo Pan Head Torx Screws 9 2 586172001 4 x 5 16 Hi Lo Pan Head Torx Screws 10 1 82533 Main Harness Assembly 11 2 60366 Nylon Cable Ties Beaded 12 2 555...

Страница 28: ...ies Scaler Service Parts Section G310 CAVITRON 300 SERIES SCALER DISPLAY FPGA PC BOARD ASSEMBLY ITEM QUANTITY PART NO DESCRIPTION 1 1 82517 Display FPGA PC Board Assembly Includes Display FPGA PC Boar...

Страница 29: ...29 3 Cavitron 300 Series Scaler Service Parts Section G310 CAVITRON 300 SERIES SCALER POWER SUPPLY MODULE ITEM QUANTITY PART NO DESCRIPTION 1 1 82529 1 Power Supply Module...

Страница 30: ...30 3 Cavitron 300 Series Scaler Service Parts Section G310 CAVITRON 300 SERIES SCALER WATER REGULATOR SOLENOID ITEM QUANTITY PART NO DESCRIPTION 1 1 629205005 Water Regulator Solenoid Assembly...

Страница 31: ...Cavitron 300 Series Scaler Service Parts Section G310 CAVITRON 300 SERIES SCALER MASTER ULTRASONIC PC BOARD WIRE HARNESS ITEM QUANTITY PART NO DESCRIPTION 1 1 82533 Master Ultrasonic PC Board Wire Har...

Страница 32: ...Series Scaler Service Parts Section G310 CAVITRON 300 SERIES SCALER INLET HARNESS WITH TOROID AND WIRE HARNESS GROUND ITEM QUANTITY PART NO DESCRIPTION 1 1 82534 Inlet Harness with Toroid 2 1 82536 Wi...

Страница 33: ...CAVITRON 300 SERIES SCALER HP CABLE BULKHEAD HARNESS USB WIRE HARNESS FPGA TO PSOC WIRE HARNESS ITEM QUANTITY PART NO DESCRIPTION 1 1 82520 Handpiece Cable Bulkhead Harness 2 1 82525 1 USB Wire Harne...

Страница 34: ...on 300 Series Scaler Service Parts Section G310 CAVITRON 300 SERIES SCALER FOOT PEDAL ASSEMBLY ITEM QUANTITY PART NO DESCRIPTION 1 1 82565 Tap On Foot Pedal Packed G1000 Includes box with Foot Pedal I...

Страница 35: ...1 Pack 1 8183203 Steri Mate 360 Handpiece Black 3 Pack 1 78688 Steri Mate Handpiece Black 1 Pack 1 78703 Steri Mate Handpiece Black 3 Pack 1 90125 Water Hose Assembly with Filter 90158 Water Filters 1...

Страница 36: ...8 10 Azabudai Minato ku Tokyo Japan 106 0041 Dentsply Sirona DeTrey GmbH Service Center Guttinger Strasse 35 D 78315 Radolfzell Germany Dentsply Sirona Korea 21 5 Teheran ro 34 gil Gangnam gu Seoul 0...

Отзывы: