Symptom/Cause
Corrective Measures
Sync cannot be activated:
1. Synchronization sequence
not performed correctly. Red
Sync button on Foot Control
was not pressed for at least 3
seconds.
2. Foot Control powered off.
1. Follow Synchronization steps carefully.
a. Press and hold Setting Icon.
b. Press Synchronization Icon.
c. Circular arrows will rotate in the lower left of the display.
d. Using a small object (ball point pen) press the red Sync button on the bottom of
the Foot Control for at least 3 seconds.
e. An audible signal will be heard indicating successfully synchronization of the Foot
Control and the scaler.
2. The Foot Control is packaged with the power OFF. The Foot Control must be
turned ON prior to use. (The Foot Control will not charge in the OFF state.)
a. The Power button must be pressed and held for 4 seconds to power ON the
Foot Control.
b. If the Foot Control power is OFF, pressing the Sync button or trying to Sync
with the Scaler will not turn the Foot Control ON.
Lock Icon cannot be activated:
1. Icon not depressed correctly.
1. Lock / Unlock Icon must be pressed and held to activate. Screen should dim when
it is locked.
Cannot adjust display
brightness:
1.
Display brightness can only
be adjusted in the Settings
screen.
1. Slide Brightness Icon up and down the scale to adjust. The brightness level
percentage should be indicated on the screen.
When not using the scaler
for a while, the screen turned
dark:
1.
Unit went into “Sleep Mode.”
1. Unit will enter “Sleep Mode” when it is not used for 1 hour. Press anywhere on the
display to resume operation.
Water Filter Icon did not
accurately indicate the
condition of the filter:
1.
Water Filter Icon illuminates
after 120 hours of actual
scaling operation.
1. The Water Filter Icon’s purpose is to remind the dental practitioner to replace the
filter on a periodic basis.