- 141 /141 -
Immunity aspects at r.f.
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
navigator should assure that it is used in such an electromagnetic environment.
Immunity test
Test level
EN 6060112
Compliance
Level
Electromagnetic environment guide
RF conducted
EN 6100046
3 Veff from
150 kHz to
80 MHz
3 Veff from
150 kHz to
80 MHz
Portable and mobile RF communications equipment should be used
no closer to any part of the device, including cables, than the
recommended separation distance calculated from that equation
applicable to the frequency of the transmitter. Recommended
separation distance
d = 1,2
!
P
from 150 kHz to 80 MHz
d = 1,2
!
P
from 80 MHz to 800 MHz
d = 2,3
!
P
from 800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m)
RF radiated
EN 6100043
3 Veff from
80 MHz to
2,5 GHz
3 Veff from
80 MHz to
2,5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Recommended working clearances between portable and mobile HF communication devices and the UNIT
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the device as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter (W)
Separation distance according to frequency
of the transmitter (m)
From 150 kHz
to 80 MHz
d = 1,2
!
P
From 80 MHz
to 800 MHz
d = 1,2
!
P
From 800 MHz
to 2 GHz
d = 2,3
!
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Notes:
(1) At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
(2) These guidelines may not apply in al situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.