7.
Daily Maintenance
Read this entire section before cleaning the unit. This product must be cleaned as
normal preparation for each patient to prevent cross infection.
The use of the sleeve is an additional precautionary measure against contamination
and does not substitute disinfection of the device.
After using, remove the sleeve. Clean the optical light guide rods and the handpiece
with cloth moisture with alcohol or cleaner. Keep solvents or flammable liquids from
the unit because they may damage its plastic housing.
Always keep the charger, handpiece and light guide well. Moisture may cause
electrical short-circuit or dangerous malfunction.
1. The curing time interval should exceed 5 seconds
for each cycle.
2. Verify the radiometer sensor which can impact the
accuracy of the measurement. The surface of sensor
area can be wiped with a cotton swab with alcohol.
3. Carefully hold the unit, so the Light Guide
Attachment is aligned with the radiometer sensor
and centered within the white circle provided.
STANDARD
05
Tips:
Optical fiber light guide must be horizontal
alignment in the middle of the sensor area.
Test the Light Guide Attachment with the Radiometer
Verify the LEDEX
TM
WL-090 performance each time before using the radiometer
which is built into the cradle.
15