Professional LED Light System for All of Surgery Application
28
8. Sterilization and Cleaning
8.1 Sterilization
WARNING
Before using MAIN HANDLE or SENSOR COVER SEALING, please conduct a
sterilization treatment.
The requirements of the national committees (standards and directives) for
disinfection must be observed.
MAIN HANDLE
CAMERA COVER OUT
SENSOR COVER SEALING
Sterilization
Ÿ
Service Condition
-
Temperature: 125 ± 5
℃
-
Pressure: 160 ± 50kPa (1.6 ±0.5 kgf/
㎠
)
-
Time: < 20 min
Autoclave
8.2 Cleaning
WARNING
Before and after use, clean the LIGHTHEAD and MAIN HANDLE with alcohol or
ethanol.
It is not allowed to use agents which contain the following substances : High-
concentration organic and inorganic acids, Chlorinated hydrocarbons.