![DentalEZ RAM VAC COSP22S Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/dentalez/ram-vac-cosp22s/ram-vac-cosp22s_user-manual_2483371023.webp)
www.DentalEZ.com
866-DTE-INFO
20
Osprey
TM
Compressor
PN: 1045DOC_E
Section IV Operation
7. The compressor installation is complete.
NOTE:
Periodically during startup from 0 PSI, the
dryer shuttle valve may be out of sync from the
previous cycle. This is noted by the sound of air
being released out of the dryer purge valves. This
minor interruption causes no harm to the dryer or
air system. The dryer shuttle valve will relocate itself
to the correct position after 35 seconds, and after
which the dryer will cycle normally.
Getting Started
(Continued)
C2 Control Display Screen
Electronic
Display
Control Buttons (4)
Remote Switch
Connection
OWL
Communication
Cable Cat 6
Motor
Disable
Figure 12. Osprey compressor C2 smart control panel
The C2 smart control platform utilizes state-of-
the-art technology with a pressure transducer
and solid state relays to control cycling of the
compressor.
Starts Unit
Stops Unit from Running
Resets Alarms and Maintenance Timer
Toggles Screens in Display
Содержание RAM VAC COSP22S
Страница 1: ...OSPREY Osprey TM Dental Compressor User Manual ...
Страница 3: ...www DentalEZ com 866 DTE INFO ii OspreyTM Compressor PN 1045DOC_E ...
Страница 13: ...www DentalEZ com 866 DTE INFO 10 OspreyTM Compressor PN 1045DOC_E ...
Страница 27: ...www DentalEZ com 866 DTE INFO 24 OspreyTM Compressor PN 1045DOC_E ...
Страница 33: ...www DentalEZ com 866 DTE INFO 30 OspreyTM Compressor PN 1045DOC_E ...
Страница 35: ...www DentalEZ com 866 DTE INFO 32 OspreyTM Compressor PN 1045DOC_E ...
Страница 37: ...www DentalEZ com 866 DTE INFO 34 OspreyTM Compressor PN 1045DOC_E ...