Diagnostics
Verification Check Troubleshooting
87
Issue
Probable Causes/ Solutions
Low Absorbance
Values
Dirty measurement surfaces or improper Blank.
Clean the surfaces, measure a new Blank using fresh source of dH
2
0
and restart the procedure using a new ampule of LC-NA.
High Absorbance
Values
Multiple measurements made using the same aliquot.
Use a fresh aliquot for each measurement.
Pipette tips not changed between aliquots.
Use a fresh tip to pipette each aliquot onto the sample surface.
The sample surfaces are not adequately cleaned between
measurements.
Wipe both the top and bottom surfaces with a dry lab wipe.
The solution has concentrated due to prolonged exposure.
Use a fresh ampule of solution.
High Standard
Deviations
Multiple measurements made using a single 1 ul aliquot.
Use a fresh aliquot for each measurement.
Pipette tips not changed between aliquots.
Use a fresh tip to pipette each aliquot onto the sample surface.
The sample surfaces are not adequately cleaned between
measurements.
Clean both the top and bottom surfaces between measurements.