8. Absorbance Colorimetrics App
Refer to the manufacturer of the assay reagent for assay specific instructions.
Quick Protocol
1. Launch the app from the
Home
screen.
2. Select the assay type on
Run
screen. Navigate to the
Standard Curve
screen.
3. Select the appropriate microvolume or cuvette mode prior to making the
Blank measurement.
4. Establish a
Blank
using the appropriate buffer or the 0.00 mg/mL standard.
5. Generate a new Standard Curve by measuring at least one replicate for a
minimum of 2 standard concentrations. Alternatively, apply a previously
saved curve selected from the drop-down menu.
6. Navigate to the
Run
screen. Enter a sample name.
• Optional: Enter a dilution factor to be applied to the reported concentration.
7. Quantitate the sample using the
Measure
button.Notes:
•
The Graph Flip button is available on the Run and Graph Screens in the
Colorimetrics and Custom Standard Curve apps. The flip option is not available
when reviewing the results in the Data app. Only the absorbance style graphs will
be displayed.
•
Standard Concentrations: Minimum 2, Maximum 8.