Denjoy DY400-4 Скачать руководство пользователя страница 8

15 

clean it with cloth. Keep it clean for better light output and curing 
performance 

5. The light guide should be sterilized after each use. It can be 

safely autoclaved at 121

, 1060hPa for 30minutes. 

6. The other accessories such as oval protector …etc., which is 

close to patients should be scrubbed by pure water or antiseptic 
and follows the standard disinfection procedure to disinfect the 
materials.  

7. Please use the original charger, any other charger may 

result in the damage of the battery and the controlled electric 
circuit; even the machine will be greatly damaged. 

 

SECTION IX. TROUBLE SHOOTING   
 

Question 1: When charging the unit, green and red light 

indicators of power charger flash by turns.   

Solution: Check whether the handpeice is connected well with 

the charging base or not. Reconnect the handpiece with the 
charging base, making sure that connection is well and stable.   

 
Question 2: The unit can not be turned on and LCD screen 

does not work. 

Solution: 

1. Possible cause: low battery. Charge with power charger at 

once for a short time, then turn on the unit again to check.   

2. Possible cause: power button   

16 

Please try to press power button 

  extendedly   

Please check the power button is in good order. 

 
Question 3: Curing program and curing time can not be 

adjusted. 

Solution: Check that function buttons are in good order. 
Question 4: Solidification effect is not good.   
Solution:  Check that the light intensity is high enough for 

curing. Check that the curing program is correct.   

 
Question 5: The unit can not be charged and the battery sign 

does not show charging condition.   

Solution:

 

 

1

Check that the light indicator of charger show red. 

2

Check that the charger or charging base is connected 

with handpiece properly.   

3

Check that the metal dome (interface) on the handpiece 

is loose or damaged.   

4) Check that the metal dome (interface) on the charging 

base is loose or damaged.   

 
Question 6: It is unable (or hard) to insert the light guide to the 

handpiece. 

Solution: Check that the size of light guide is suitable for 

handpiece.  

Содержание DY400-4

Страница 1: ......

Страница 2: ...ign Beep at every 5 seconds to remind the user of remaining curing time 3 Cord Cordless interchangeable Cordless or cord use are both available When running out of battery the unit can be still used d...

Страница 3: ...long time It can guarantee the stability of performance to avoid over heat Not many manufacturers make the unit with radiating fan inside 10 Build in meter on the front of charging base to check the...

Страница 4: ...ing Program 6 Charger Port 7 curing Switch 8 Charging interface 9 Charger Port 10 Light Intensity Indicator 11 Status Indicator 12 Test Hole 13 Charger Connector Standard Accessories Handpiece 1 PC Ch...

Страница 5: ...g it shows that the battery is fully charged And the light indicator of charger turns into green accordingly 10 2 Install the eye protection cover Directly insert it onto the front of the handpiece It...

Страница 6: ...charger with the handpiece for continuous use 12 6 Turn off the power When the unit turns into sleeping mode automatically it will shut down all the function except the battery sign beep sound sign an...

Страница 7: ...at the light emitted from this curing unit Do not use this device without suitable 14 protective eye ware for the operator assistant and patient Suitable protective eye ware blocks all energy below 55...

Страница 8: ...be turned on and LCD screen does not work Solution 1 Possible cause low battery Charge with power charger at once for a short time then turn on the unit again to check 2 Possible cause power button 1...

Страница 9: ...and confronts to the current products specification This instrument is guaranteed for its designated use against original defects in materials and workmanship for a period of 12 months from date of pu...

Отзывы: