The 

reading comes from the meter is as reference for 

dentist to adjust the appropriate curing time. Dentists still need to 
follow the curing procedures which are provided by composite 
manufacturer.  

 

The light source must aim at the fixed position when 

operated. The working distance must be within 3mm from the 
fixed position. The depth of the polymerization of material is not 
less than 2mm. Curing layer by layer is recommended if the 
material is very thick!   

 
 

SECTION VII. SAFETYPRECAUTIONS 
 

a). Before operation, you have to read usage manual 

carefully. 

b). 

 CAUTION: This curing unit produces high output 

curing energy!   

Do not place light directly on or toward unprotected gingival 

or skin. 

he light source has to shoot directly in the polymerization 

of resin material when it is in clinical use. Improper radiation 
which may affect the curing result should be avoided. 

 c).  

  CAUTION: Do no look directly at the light emitted 

from this curing unit. Do not use this device without suitable 

14 

protective eye ware for the operator, assistant and patient. 
Suitable protective eye ware blocks all energy below 550nm. 

d). 

 

CAUTION: Persons having a history of 

photosensitive reactions or who are using photosensitizing drugs 
should not be exposed to light from this unit. 

e).  

 CAUTION: Equipment not suitable for use in the 

presence of flammable anesthetic mixture with air or nitrous 
oxide. 

f). To keep the safety operation, we suggest that check your 

local AC power supply voltage before you buy this product 
oversea. 

g) The curing light should be placed in the original packing 

box in a dry and clean cupboard, in case of its drop onto the 
floor. 

 

SECTION VIII. MAINTENANCE     
 

1. This unit can not be dismantled privately; otherwise the unit 

will be damaged wholly.     

2. For non-use for a long time, the unit should be cleaned and 

then put away it inside the original package.   

3. Be sure there is no any resin or material sticking on the 

surface of the light guide 

4. If the materials are sticking on the light guide, immediately 

Содержание DY400-4

Страница 1: ......

Страница 2: ...ign Beep at every 5 seconds to remind the user of remaining curing time 3 Cord Cordless interchangeable Cordless or cord use are both available When running out of battery the unit can be still used d...

Страница 3: ...long time It can guarantee the stability of performance to avoid over heat Not many manufacturers make the unit with radiating fan inside 10 Build in meter on the front of charging base to check the...

Страница 4: ...ing Program 6 Charger Port 7 curing Switch 8 Charging interface 9 Charger Port 10 Light Intensity Indicator 11 Status Indicator 12 Test Hole 13 Charger Connector Standard Accessories Handpiece 1 PC Ch...

Страница 5: ...g it shows that the battery is fully charged And the light indicator of charger turns into green accordingly 10 2 Install the eye protection cover Directly insert it onto the front of the handpiece It...

Страница 6: ...charger with the handpiece for continuous use 12 6 Turn off the power When the unit turns into sleeping mode automatically it will shut down all the function except the battery sign beep sound sign an...

Страница 7: ...at the light emitted from this curing unit Do not use this device without suitable 14 protective eye ware for the operator assistant and patient Suitable protective eye ware blocks all energy below 55...

Страница 8: ...be turned on and LCD screen does not work Solution 1 Possible cause low battery Charge with power charger at once for a short time then turn on the unit again to check 2 Possible cause power button 1...

Страница 9: ...and confronts to the current products specification This instrument is guaranteed for its designated use against original defects in materials and workmanship for a period of 12 months from date of pu...

Отзывы: