Denjoy DY400-4 Скачать руководство пользователя страница 5

SECTION IV. MAIN TECHNICAL INDEX 

1. Classification: Class I   
2. Power source: 
a) Power supply-Power Charger:   

Input: 100-240V, 50-60Hz   Output voltage: 8.7-8.8V DC               
Output current: 500mA       

b) Internal power supply- Lithium battery 
Voltage: 7.4V                          Capacity: 1350mAh 
Charging time:    Approx. 2.5 hours (New Battery for 4 hours)   
Continuous Operating time:    Avg. 40 min with fully charged 

battery 

3. LED lamp   
Model: 5W,       Wavelength: 420-480nm 
4. Dimension: W300×H120×D150mm       
5. Net Weight: about 700g   

 

                                     

SECTION V. INSTALLATION   

1. Charge battery   

For best performance—charging time is at least 2.5 hours! 

(Charging time is at least 4 hours for the first time). The light 
indicator of charger is red when charging,   

When full battery sign 

on the LCD panel appear and 

stop flashing, it shows that the battery is fully charged. And the 
light indicator of charger turns into green accordingly.   

10 

2. Install the eye-protection cover   
Directly insert it onto the front of the handpiece. It still can be 

rotated when proper installed. 

 

3. Install light guide   

Directly insert the light guide to the handpiece, make sure it 

has been totally insert and fixed steadily. It still can be rotated 
when proper installed. 

After use, the user is free to pull out the light guide with care 

and keep it in the package.   

 

Optional

 

4. Connect power charger 

    This unit is available to use directly with power charger when 

battery run out of power. 

    Directly connect the power charger to the connector which 

is located on the bottom of handpiece, then plug the charger 
into the outlet. (AC100-240V). 

 

SECTION VI. OPERATION   

  Power switch and function switch   

1. Turn on the unit.   

Extended press the power button can turn on the unit. LCD 

panel is working. 

 

Содержание DY400-4

Страница 1: ......

Страница 2: ...ign Beep at every 5 seconds to remind the user of remaining curing time 3 Cord Cordless interchangeable Cordless or cord use are both available When running out of battery the unit can be still used d...

Страница 3: ...long time It can guarantee the stability of performance to avoid over heat Not many manufacturers make the unit with radiating fan inside 10 Build in meter on the front of charging base to check the...

Страница 4: ...ing Program 6 Charger Port 7 curing Switch 8 Charging interface 9 Charger Port 10 Light Intensity Indicator 11 Status Indicator 12 Test Hole 13 Charger Connector Standard Accessories Handpiece 1 PC Ch...

Страница 5: ...g it shows that the battery is fully charged And the light indicator of charger turns into green accordingly 10 2 Install the eye protection cover Directly insert it onto the front of the handpiece It...

Страница 6: ...charger with the handpiece for continuous use 12 6 Turn off the power When the unit turns into sleeping mode automatically it will shut down all the function except the battery sign beep sound sign an...

Страница 7: ...at the light emitted from this curing unit Do not use this device without suitable 14 protective eye ware for the operator assistant and patient Suitable protective eye ware blocks all energy below 55...

Страница 8: ...be turned on and LCD screen does not work Solution 1 Possible cause low battery Charge with power charger at once for a short time then turn on the unit again to check 2 Possible cause power button 1...

Страница 9: ...and confronts to the current products specification This instrument is guaranteed for its designated use against original defects in materials and workmanship for a period of 12 months from date of pu...

Отзывы: