54
Manufacturer’s manual and declaration – interference resistance
These devices are intended for use in the electromagnetic environment specified below. The customer or the user of
the device should ensure its use in the specified electromagnetic environment.
Interference
resistance test
The test level
according to IEC 60601
Compliance
level
Electromagnetic environment – instructions
Conducted
disturbances
induced by RF fields
according to
IEC 61000-4-6
Radio-frequency
electromagnetic field
according to
IEC 61000-4-3
3 V (150 kHz - 80 MHz)
6 V (ISM & Amateur)
10 V/m
(80 MHz - 2.7 GHz)
Complies
Complies
The distance between the mobile radiotelephone
communica on systems used and any element of
the device, including the cables, should not be less
than the recommended separa on distance which
is calculated in accordance with the expressions
below with reference to the frequency of the trans-
mi er.
The field density in the propagation of radio waves
from stationary radio transmitters, according to
the results of observations of the electromagnetic
situation, should be lower than the level of
correspondence in each frequency band.
The effect of interference may occur near the
equipment marked with the symbol
Table 3