User Manual for PC-80D Easy ECG Monitor
27
Appendix 1 Additional measuring method
Direction of using limb clip type electrodes: clean the electrodes by using wet cloth; plug the lead wires into socket; or
affix the lead wires to clip, and then keep the clip in measuring position.
Direction of using sucking bulb electrode: Squeeze the rubber bulb with your hand to make the metal dish place on the
measuring position.
1. Limb lead measurement (
paste on chest
)
This measurement applies to long term measurement for ECG signal of limb lead I, II, or III. The electrodes are adhesive
ECG electrodes.
The location of the electrodes: (as shown in figure 1)
R: below the clavicle and close to right shoulder;
L: below the clavicle and close to left shoulder;
N: on the right lower abdomen;
F: on the left lower abdomen.
2.
Chest lead measurement (
paste on chest
)
R (red): below the clavicle and close to left shoulder;
L (yellow): below the clavicle and close to right shoulder;
F (green): on chest, please refer to the figure 4 (location of V1-V6 is optional).
N (black): unused.
3.
Chest lead measurement (
limb clip type electrodes
+
sucking bulb electrode
)
R(red): place on interior of left arm (between wrist and elbow)
;
L (yellow): place on interior of the right arm (between wrist and elbow);
F sucking bulb electrode (green): on chest
,
please refer to the figure 4 -1(location of V1-V6 is optional)
;
N (black): unused.
Figure 2 chest lead (paste on chest)
Figure 3 chest lead
(
limb clip type electrodes
+
sucking bulb electrode
)
Figure 1 limb lead (paste on chest)