User Manual for PC-80D Easy ECG Monitor
10
(9)
“
”:
Status bar of measurement progress.
Operation:
When in the quick measurement by built-in electrodes:
①
Press “
” key, the measuring display screen will be made a quarter turn counterclockwise;
②
Press the “
” ok key or “
”
return key to stop measuring and return to the default menu screen directly, the measuring data won’t be saved.
When in the external electrodes lead wire measurement:
①
The device will stop measuring automatically if the measuring time is over;
②
Press the “
” ok key or “
”
return
key to stop measuring and return to the default menu screen directly, the measuring data would be saved.
Note:
①
If the storage is full, the device will stop to measure and it prompts “Fail to save this record because of full data memory”, at this moment, you
should delete some records to make sure enough storage space, then measuring again.
②
If the battery is low, the battery icon “
” flashes red, the device stop to measure and it prompts “low battery voltage, measuring is interrupted,
return key to exit! ”, at this time, connect to external power can continue to measure, or exchanging a new battery to measure again.
③
During measuring mode, it will have regular pulse beep sound.
④
During the measurement by external lead wires, if “HR alarm” option is set to “On” and the HR value is out of the alarm limits, the alarm sounds
will occur with 3 short beeps, and the HR value is reversely displayed.
⑤
During the quick measurement, if the device does not check well contact with the electrodes or the signal is poor, it will not be calculate the
average HR value, the measuring result is “poor signal, measure again”, this data will not be saved.
⑥
During the measurement, if the placement of electrodes loose contact with the electrodes, it will prompt “poor signal”. If it is external lead wire
measurement, the HR value shows “__”.