Chapter 4 Communication Setting
LD-E Operation Manual
4-2
0x0E: ASCII, 7, O, 2
0x0F: ASCII, 7, E, 1
0x10: ASCII, 7, E, 2
1004H~1005H (R)
(03H)
Absolute position
Displacement absolute position output:
LD-100E: 65000~135000um
LD-220E: 120000~320000um
LD-500E: 300000~700000um
1006H~1007H (R)
(03H)
Relative position
Relative position of displacement = Absolute
position - Zero position
1008H (R/W)
(03H,06H)
Set the zero value
0x00: Turn off zero function
、
0x01: Set the
zero
1009H~ 100AH (R)
(03H)
Zero position
Zero position
100BH (R/W)
(03H,06H)
One_SL position
Comparator lower limit position (unit:0.1mm)
100CH
Reserved
100DH (R/W)
(03H,06H)
Two_SL position
Comparator upper limit position (unit:0.1mm)
100EH
Reserved
100FH (R/W)
(03H)
Weight position
Weight position on PD
1010H (R)
(03H)
Exposure class
Exposure level in auto exposure mode
(1~100)
1011H (R/W)
(03H,06H)
Reaction speed
0x01: 1.5ms (Default)
0x03: 3ms
0x05: 5ms
1012H (R/W)
(03H,06H)
Moving average
setting
0x01: None
0x02:
2 times on average
0x04: 4 times on average
0x08: 8 times on average
0x10: 16 times on average
0x20: 32 times on average
0x40: 64 times on average
0x80: 128 times on average (Default)
0x100: 256 times on average
0x200: 512 times on average
1013H (R/W)
(03H,06H)
Median filter
setting
0x01: Window size 1
0x03: Window size 3
0x05: Window size 5
0x07: Window size 7 (Default)
0x09: Window size 9
0x0B: Window size 11
0x0D: Window size 13
Содержание LD-100E-M22
Страница 4: ...Chapter 1 Product Profile LD E Operation Manual 1 3 Wiring Instructions...
Страница 14: ...Chapter 2 Settings and Measurements LD E Operation Manual 2 9 Steps...
Страница 15: ...Chapter 2 Settings and Measurements LD E Operation Manual 2 10...
Страница 16: ...Chapter 2 Settings and Measurements LD E Operation Manual 2 11...
Страница 19: ...Chapter 3 Software Operation LD E Operation Manual 3 2 Next Close...
Страница 25: ...Chapter 3 Software Operation LD E Operation Manual 3 8 Parameter reading and writing operation area...