Technical Reference
87
HH2 displays ‘Sys Error’.
The most likely cause of ‘Sys Error’ is corrupted calibration values or other
configuration settings. HH2 firmware (PROM) Versions earlier than 1.08 were
vulnerable to corruption of calibration values if the battery was not connected cleanly.
Solution: Cold Boot the HH2.
See To Cold Boot the HH2 on page86
Upgrade your HH2 to v2 firmware as soon as possible. Contact Delta-T.
Readings Corrupt,
Store? prompt missing
On initial power up the HH2 reports a sequence of messages such as Readings
Corrupt OR, after taking a reading, the HH2 fails to display the "Store?" prompt and it
is not possible to record readings.
To clear the error condition cold-boot the HH2 but note that this will destroy any data
that are already stored. So try to download any stored readings before proceeding
(no guarantee of success if readings are corrupted).
See To Cold Boot the HH2 on page86
The HH2 displays the error condition only when it first detects the error, normally on
power-up, e.g. "Readings Corrupt" messages that occur when it fails to cold-boot
correctly. It then patches up the error where possible, for example, restoring the
default linearisation tables from ROM, and resumes quasi-normal operation. This
means that it is able to take readings and doesn't report the error any more.
Error code 7 – on HH2Read
This problem is common when the RS232 cable is pulled out of the HH2 half way
through downloading readings or because the HH2 has detected an error condition
with some of the readings. Unfortunately if this has happened the readings can no
longer be accessed and the only way to restore the HH2 will be to cold boot it.
Note: readings will be lost.
See To Cold Boot the HH2 on page86
Display is Blank
The meter will sleep when not used for more than 30 seconds. This means the
display will go blank.
First check that the meter is not sleeping by pressing the Esc key. The display
should become visible instantly.
If the display remains blank, then try all the keys in case one key is faulty.
If you have a voltmeter, then check the battery exceeds 5 Volts. If it drops below this
value then the performance of the meter can be intermittent. If it is a long way below
5 Volts then the meter may not work at all.
Try replacing the battery.
If you are in bright light, then the display may be obscured by the light shining on the
display. Try to move to a darker area or shade the display.
Содержание HH2
Страница 1: ...User Manual for the Moisture Meter type HH2 Delta T Devices Ltd Version 4 0 1 ...
Страница 17: ...10 Navigating Options and Readings Moisture Meter User Manual 4 0 ...
Страница 25: ...18 SM200 and ThetaProbes Moisture Meter User Manual 4 0 ...
Страница 35: ...28 Profile Probes Moisture Meter User Manual 4 0 ...
Страница 45: ...38 Profile Probes Moisture Meter User Manual 4 0 ...
Страница 57: ...50 WET Sensor Moisture Meter User Manual 4 0 ...
Страница 59: ...52 Equitensiometers Moisture Meter User Manual 4 0 ...
Страница 92: ...Technical Reference 85 EMC Certificate ...