HD45 / HD46
- 61 -
V2.4
Dewpoint Td accuracy (°C)
The dewpoint is a computed quantity that depends on relative humidity and
temperature calibration accuracy. The values supplied below refer to
±
2.5%RH,
±
0.25°C, 1013.25mbar accuracies.
Relative humidity (%)
10
30
50
70
90
100
-20
2.50 1.00 0.71 0.58 -- --
0
2.84 1.11 0.78 0.64 0.56
0.50
20
3.34 1.32 0.92 0.75 0.64
0.62
50
4.16 1.64 1.12 0.90 0.77
0.74
Tempera
tu
re °
C
)
100
5.28 2.07 1.42 1.13 0.97
0.91
Instrument features
Measuring frequency
1 sample every 3 seconds
Memory size
2304 Records
Logging interval
Selectable between: 30s, 1m and 5m
The stored values represent the average values of the
samples acquired every 3 seconds in the selected logging
interval.
Serial outputs
mini-USB (mini-USB / USB cable RS45 or RS45I cod.)
RS485 MODBUS-RTU
(HD45…S… and HD46…S… only)
Stored data security
Unlimited
Analog outputs
0…10Vdc (R
L
> 10k
Ω
)
(HD45…V… and HD46…V only)
11Vdc out of range
4…20mA (R
L MAX
= 400
Ω
)
(HD45…A… and HD46…A only)
22mA outside the measuring range
Active sourcing current output
Relay outputs
Bistable relay
(HD45…R and HD46…R only)
Contact: max 1A @ 30Vdc resistive load
Power supply
24Vac
±
10% (50…60Hz) or 15…35Vdc
Power consumption
100 mW (
except models with current output
)
400 mW (
for models with current output
)
Stabilization time at switch-on
15 minutes (to ensure declared accuracy)
Instrument operating
temperature
0°C ... 50°C
Operating relative humidity
0%RH ... 95%RH non-condensing
Dimensions
(LxHxW)
80 x 80 x 30 mm
(HD45.17…)
80 x 80 x 34 mm
(HD45.B… and HD45.7B…)
120 x 80 x 30 mm
(HD46.17…)
120 x 80 x 34 mm
(HD46.17B…)
Weight
50 g
Housing material
ABS