HD37AB17D/HD37B17D 

-   8   -

 V1.5

 

3.1.1. Batteries recharge 

To recharge the batteries package, use the 

SWD06 

battery charger provided with the 

instrument. 

Go on as indicated below: 

 

connect the pin of the battery charger to the power supply net and the connector 
of the battery charger to the pin placed at the left of the instrument.  

The power 

supply must be 6Vdc

 

If the batteries are particularly discharged or their temperature is about zero, the 
led CHARGE flashes for some seconds (pre-charge phase) and then it remains fix-
edly on in order to indicate that the instrument is in the speed charge phase. 

 

Charge the batteries till the led CHARGE doesn't switch off. 

 

3.1.2. Notes for the batteries use 

 

At the first start up, it’s necessary to completely recharge the batteries. 

 

The charge time of the batteries package is about 4 hours. 

 

The last of the batteries package in measurement working mode is about 8 hours. 

 

A new Ni-MH batteries package reaches the maximum of its performance only after 

being discharged and charged completely again at least twice or three times. 

 

The batteries package autonomy depends on the instrument use. Even if the in-
strument is in stand-by with the batteries package completely charges, it is 
autonomously charged during the time. 

 

The batteries package can be charge and discharge hundred of times, but using 
them the charge loses its own capacity. Replace the batteries package when the 
autonomy is reduced at some hours. 

 

Use only Delta Ohm batteries package code 

BAT-20 

and recharge it using

 SWD06 

battery charger or one that complies with the specifications indicated in the techni-
cal data.

 

 

The Ni-MH batteries package lasts more if, sometimes, you act with cutting and 
you completely discharge it. 

 

Extreme temperatures weigh negatively on the performances of the batteries 
package. 

 

At the beginning of a new charge, you can check a variation of the internal tem-
perature of the instrument that can temporarily influence the CO

measurement. 

Содержание HD37AB17D

Страница 1: ...Operating manual Data logger Temperature RH CO CO2 HD37AB17D HD37B17D www deltaohm com Companies Brands of GHM English Keep for future reference...

Страница 2: ...HD37AB17D HD37B17D 2 V1 5 Datalogger RH Temperature CO CO2 HD37AB17D and HD37B17D 1 2 3 5 4 6 7...

Страница 3: ...7AB17D 3 It visualizes the working state of the instrument 4 It offers the information about the charge state of the Ni MH rechargeable bat teries 5 Button ON OFF it switches the instrument ON and OFF...

Страница 4: ...arge 8 3 1 2 Notes for the batteries use 8 3 1 3 Replacement of the batteries package 9 3 1 4 Batteries disposal 9 3 2 CONNECTION TO THE PC 10 3 2 1 Serial commands 10 4 SENSORS CALIBRATION 12 4 1 RH...

Страница 5: ...on sensor of NTC type The sensor for CO Carbon Monoxide only HD37AB17D measurement consists of an electrochemical cell with two electrodes suitable to detect the presence of carbon monoxide lethal for...

Страница 6: ...ngs so release the button To switch the instrument off stand by mode press the button ON OFF till the Led Activity starts emitting two long and close flashings so release the button It s not possible...

Страница 7: ...it feeds the instrument and re charges the Ni MH batteries package Power supply connector The power supply connector has a 5 5mm external diameter and a 2 1mm internal diameter The positive pole is t...

Страница 8: ...mode is about 8 hours A new Ni MH batteries package reaches the maximum of its performance only after being discharged and charged completely again at least twice or three times The batteries package...

Страница 9: ...nt unfastening the screw 1 Take off the connector paying attention not to tear the wires away Remove the batteries package Connect the connector to the new batteries package the connector has a refer...

Страница 10: ...of a driver inserted into the DeltaLog13 software CD ROM Before connecting the USB cable to the PC in stall the driver please see the documentation provided with DeltaLog13 CD ROM The parameters of t...

Страница 11: ...A00ZKE It resets the number of the sec tions and of the records A00ZKZ It resets the Flag of Automatic Start A00ZRL 00000 Length in seconds of the pro grammed Log A00ZRM 06800 Next record number in lo...

Страница 12: ...alibrating This calibration process cancels the data of the previous calibrations For a correct sensor calibration the first point must be at 75 RH and the second one at 33 RH 1 Connect the instrument...

Страница 13: ...on and wait at least 15 minutes because the measurement must stabilize 2 Connect the instrument to the PC by USB connection cable code CP22 and start DeltaLog13 program so follow the instruction conn...

Страница 14: ...If it is necessary to replace the CO sensor proceed as follows 1 Switch the instrument off 2 Using the screwdriver open the wicket of the sensor holder and pull the flat CO sensor out 3 Take the new C...

Страница 15: ...screw the metal tube that is placed at the end of the plastic one connecting to the MINICAN 12A cylinder for calibration 3 If you want to calibrate around 400ppm let the inlet open in this second cal...

Страница 16: ...Baudrate 460800 Batteries charger power supply code SWD06 2 pole connector positive at the centre Output voltage 6Vdc Maximum current 1600mA 9 60VA Max Measuring rate 1 sample every three seconds Stor...

Страница 17: ...ectro chemical cell Measurement range 0 500ppm Sensor working range 5 50 C Accuracy 3ppm 3 of the measurement Resolution 1ppm Response time T90 50 sec Long term stability 5 of the measurement year Exp...

Страница 18: ...ated temperature sensor P6 10 m sintered AISI 316 protection for probes 14 mm thread M12 1 P7 20 m PTFE protection for probes 14 mm thread M12 1 P8 20 stainless steel grid and Pocan protection for pro...

Страница 19: ......

Страница 20: ...ird parties is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic equip...

Страница 21: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Страница 22: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V1 5 13 11 2017...

Отзывы: