1. INTRODUCTION 

 

HD37AB17D

 and 

HD37B17D 

instruments are

 dataloggers 

able to simultaneously 

measure and memorize the following parameters: 

 

Relative Humidity 

RH

 

 

Ambient temperature 

T

 

 

Carbon Monoxide 

CO

 (

only HD37AB17D

 

Carbon dioxide 

CO

2

 

HD37AB17D

 and 

HD37B17D 

are instruments suitable to research and monitor the 

air quality in internal environments.

 

Typical applications are the examination of the air quality in all the buildings where 

there is a crowd of people (schools, hospitals, auditoria, canteens, etc.), in the work 
environments in order to optimize the comfort and in the buildings in order to check if 
there is a small emission of CO, with the risk of explosions or fires. This analysis al-

lows setting the air-conditioning unit (temperature and humidity) and ventilation (air 
replacement/hour) in order to get a double purpose: getting a good quality of the air 
according to ASHRAE and IMC regulations in force and an energetic saving. 

HD37AB17D

 and 

HD37B17D

 are instruments suitable to fight the so-called syn-

drome of the ill building.

 

RH 

measurement (Relative Humidity) is obtained with a sensor of capacitive type. 

temperature is measured with a high-precision sensor of NTC type

.

 

The sensor for 

CO

 (Carbon Monoxide, 

only HD37AB17D

) measurement consists of 

an electrochemical cell with two electrodes suitable to detect the presence of carbon 
monoxide, lethal for man, in residential and industrial environments. 

CO

(Carbon Dioxide) measurement is obtained with the special infrared sensor 

(NDIR technology: Non-Dispersive Infrared Technology) that, thanks to the use of a 
double filter and a special measurement technology, warranties precise and constant 

measurements over the time. The presence of a protection membrane through which 
is given off the air to check, protects the sensor from dust and atmospheric agents.  

HD37AB17D

 and 

HD37B17D 

are 

dataloggers

 able to memorize the detected 

measurements, with a frequency prearranged by the user.

 

HD37AB17D

 and 

HD37B17D 

connect to the PC through

 USB 

inlet.

 

The instruments are supplied with 

DeltaLog13 Software

 with which you can manage 

the connection operations to the PC, the RH, CO (only HD37B17D) and CO

sensor 

calibrations, the instrument working parameters setup, the data transfer, the graphi-
cal introduction and the acquired or memorized measurements print. 
 

Through an appropriate procedure

the Software 

DeltaLog13

 can evaluate the pa-

rameter 

% OA 

(percentage of external air), according to the formula: 

 

100

%

=

o

r

s

r

X

X

X

X

OA

 

whereas:  

X

=  

CO

in return air 

 

X

s

=  

CO

in the outlet air

 

X

0

=  

CO

2  

in the external air 

Содержание HD37AB17D

Страница 1: ...Operating manual Data logger Temperature RH CO CO2 HD37AB17D HD37B17D www deltaohm com Companies Brands of GHM English Keep for future reference...

Страница 2: ...HD37AB17D HD37B17D 2 V1 5 Datalogger RH Temperature CO CO2 HD37AB17D and HD37B17D 1 2 3 5 4 6 7...

Страница 3: ...7AB17D 3 It visualizes the working state of the instrument 4 It offers the information about the charge state of the Ni MH rechargeable bat teries 5 Button ON OFF it switches the instrument ON and OFF...

Страница 4: ...arge 8 3 1 2 Notes for the batteries use 8 3 1 3 Replacement of the batteries package 9 3 1 4 Batteries disposal 9 3 2 CONNECTION TO THE PC 10 3 2 1 Serial commands 10 4 SENSORS CALIBRATION 12 4 1 RH...

Страница 5: ...on sensor of NTC type The sensor for CO Carbon Monoxide only HD37AB17D measurement consists of an electrochemical cell with two electrodes suitable to detect the presence of carbon monoxide lethal for...

Страница 6: ...ngs so release the button To switch the instrument off stand by mode press the button ON OFF till the Led Activity starts emitting two long and close flashings so release the button It s not possible...

Страница 7: ...it feeds the instrument and re charges the Ni MH batteries package Power supply connector The power supply connector has a 5 5mm external diameter and a 2 1mm internal diameter The positive pole is t...

Страница 8: ...mode is about 8 hours A new Ni MH batteries package reaches the maximum of its performance only after being discharged and charged completely again at least twice or three times The batteries package...

Страница 9: ...nt unfastening the screw 1 Take off the connector paying attention not to tear the wires away Remove the batteries package Connect the connector to the new batteries package the connector has a refer...

Страница 10: ...of a driver inserted into the DeltaLog13 software CD ROM Before connecting the USB cable to the PC in stall the driver please see the documentation provided with DeltaLog13 CD ROM The parameters of t...

Страница 11: ...A00ZKE It resets the number of the sec tions and of the records A00ZKZ It resets the Flag of Automatic Start A00ZRL 00000 Length in seconds of the pro grammed Log A00ZRM 06800 Next record number in lo...

Страница 12: ...alibrating This calibration process cancels the data of the previous calibrations For a correct sensor calibration the first point must be at 75 RH and the second one at 33 RH 1 Connect the instrument...

Страница 13: ...on and wait at least 15 minutes because the measurement must stabilize 2 Connect the instrument to the PC by USB connection cable code CP22 and start DeltaLog13 program so follow the instruction conn...

Страница 14: ...If it is necessary to replace the CO sensor proceed as follows 1 Switch the instrument off 2 Using the screwdriver open the wicket of the sensor holder and pull the flat CO sensor out 3 Take the new C...

Страница 15: ...screw the metal tube that is placed at the end of the plastic one connecting to the MINICAN 12A cylinder for calibration 3 If you want to calibrate around 400ppm let the inlet open in this second cal...

Страница 16: ...Baudrate 460800 Batteries charger power supply code SWD06 2 pole connector positive at the centre Output voltage 6Vdc Maximum current 1600mA 9 60VA Max Measuring rate 1 sample every three seconds Stor...

Страница 17: ...ectro chemical cell Measurement range 0 500ppm Sensor working range 5 50 C Accuracy 3ppm 3 of the measurement Resolution 1ppm Response time T90 50 sec Long term stability 5 of the measurement year Exp...

Страница 18: ...ated temperature sensor P6 10 m sintered AISI 316 protection for probes 14 mm thread M12 1 P7 20 m PTFE protection for probes 14 mm thread M12 1 P8 20 stainless steel grid and Pocan protection for pro...

Страница 19: ......

Страница 20: ...ird parties is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic equip...

Страница 21: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Страница 22: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V1 5 13 11 2017...

Отзывы: