3.2. CONNECTION TO THE PC 

The instrument is provided with a 2.0 USB communication interface B type. For the 
connection to the PC, it’s required the cable code 

CP22. 

The instruments are supplied with 

DeltaLog13 software. 

With the software you can 

manage the connection operations to the PC, the CO

2

 sensor calibration (only 

HD37B17D), the RH calibrations, the instrument working parameters setup, the data 
transfer, the graphical introduction, the acquired or memorized measurements print. 

DeltaLog13 software is endowed with a “Help on-line” 

(also in pdf format) that 

describes the features and the functions.

 

Moreover, the instruments are compatible with HyperTerminal communication pro-

gram supplied together with Windows operational systems (from Windows 98 to Win-
dows XP). 
 

3.2.1 Serial commands 

The connection through USB requires the prior installation of a driver inserted into the 

DeltaLog13

 software CD-ROM. 

Before connecting the USB cable to the PC, in-

stall the driver (please see the documentation provided with DeltaLog13 CD-

ROM). 

The parameters of the instrument USB serial transmission are: 

 

Baud rate  480600 baud 

 

Parity None 

 

Bit number  8 

 

Bit stop 

 

Protocol  Xon / Xoff. 

The instruments are provided with a full kit of controls and request of data sent by the 
PC. 
All the controls sent to the instrument must have the following structure: 

A00ZXXCR 

where: 

XX is the control code and CR is the Carriage Return (code 

ASCII 0D). 

The XX control characters are only capitals, the instrument responses with “&” if the 
key is right, with  “?” if any combination of characters is wrong. 
The response strings of the instrument finish with 

“|” 

character (code

 

ASCII  124) fol-

lowed by the enter of the keys (Carriage Return) and LF (Line Feed). 

  

Key 

Response

 

Description

 

A00ZP0 &| 

Ping 

A00ZS0  

 

A00ZS1  

 

A00ZS2  

 

A00ZDAaammgghhmmss  

 

A00ZDSaammgghhmmss  

 

A00ZDLxxxxx  

 

A00ZFA  

 

A00ZFS  

 

A00ZG0 

& HD37AB17D | 

Instrument code 

A00ZG1 

& V1R1 2008/12/05| 

Version and date of the Firmware 

Содержание HD37AB17D

Страница 1: ...Operating manual Data logger Temperature RH CO CO2 HD37AB17D HD37B17D www deltaohm com Companies Brands of GHM English Keep for future reference...

Страница 2: ...HD37AB17D HD37B17D 2 V1 5 Datalogger RH Temperature CO CO2 HD37AB17D and HD37B17D 1 2 3 5 4 6 7...

Страница 3: ...7AB17D 3 It visualizes the working state of the instrument 4 It offers the information about the charge state of the Ni MH rechargeable bat teries 5 Button ON OFF it switches the instrument ON and OFF...

Страница 4: ...arge 8 3 1 2 Notes for the batteries use 8 3 1 3 Replacement of the batteries package 9 3 1 4 Batteries disposal 9 3 2 CONNECTION TO THE PC 10 3 2 1 Serial commands 10 4 SENSORS CALIBRATION 12 4 1 RH...

Страница 5: ...on sensor of NTC type The sensor for CO Carbon Monoxide only HD37AB17D measurement consists of an electrochemical cell with two electrodes suitable to detect the presence of carbon monoxide lethal for...

Страница 6: ...ngs so release the button To switch the instrument off stand by mode press the button ON OFF till the Led Activity starts emitting two long and close flashings so release the button It s not possible...

Страница 7: ...it feeds the instrument and re charges the Ni MH batteries package Power supply connector The power supply connector has a 5 5mm external diameter and a 2 1mm internal diameter The positive pole is t...

Страница 8: ...mode is about 8 hours A new Ni MH batteries package reaches the maximum of its performance only after being discharged and charged completely again at least twice or three times The batteries package...

Страница 9: ...nt unfastening the screw 1 Take off the connector paying attention not to tear the wires away Remove the batteries package Connect the connector to the new batteries package the connector has a refer...

Страница 10: ...of a driver inserted into the DeltaLog13 software CD ROM Before connecting the USB cable to the PC in stall the driver please see the documentation provided with DeltaLog13 CD ROM The parameters of t...

Страница 11: ...A00ZKE It resets the number of the sec tions and of the records A00ZKZ It resets the Flag of Automatic Start A00ZRL 00000 Length in seconds of the pro grammed Log A00ZRM 06800 Next record number in lo...

Страница 12: ...alibrating This calibration process cancels the data of the previous calibrations For a correct sensor calibration the first point must be at 75 RH and the second one at 33 RH 1 Connect the instrument...

Страница 13: ...on and wait at least 15 minutes because the measurement must stabilize 2 Connect the instrument to the PC by USB connection cable code CP22 and start DeltaLog13 program so follow the instruction conn...

Страница 14: ...If it is necessary to replace the CO sensor proceed as follows 1 Switch the instrument off 2 Using the screwdriver open the wicket of the sensor holder and pull the flat CO sensor out 3 Take the new C...

Страница 15: ...screw the metal tube that is placed at the end of the plastic one connecting to the MINICAN 12A cylinder for calibration 3 If you want to calibrate around 400ppm let the inlet open in this second cal...

Страница 16: ...Baudrate 460800 Batteries charger power supply code SWD06 2 pole connector positive at the centre Output voltage 6Vdc Maximum current 1600mA 9 60VA Max Measuring rate 1 sample every three seconds Stor...

Страница 17: ...ectro chemical cell Measurement range 0 500ppm Sensor working range 5 50 C Accuracy 3ppm 3 of the measurement Resolution 1ppm Response time T90 50 sec Long term stability 5 of the measurement year Exp...

Страница 18: ...ated temperature sensor P6 10 m sintered AISI 316 protection for probes 14 mm thread M12 1 P7 20 m PTFE protection for probes 14 mm thread M12 1 P8 20 stainless steel grid and Pocan protection for pro...

Страница 19: ......

Страница 20: ...ird parties is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic equip...

Страница 21: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Страница 22: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V1 5 13 11 2017...

Отзывы: