16
MAN-039-1-1
Specifications
Physical Specifications
Dimension (Body)
27 x 46 x 13 mm
Dimension (Head)
23 x 30 x 7 mm
Cable Length “Quattro”
254 mm, 229 mm, 203 mm, 178 mm
Cable Length “Duo”
254 mm
Cable Length “Mini”
203 mm
Mass (“Quattro”, “Duo”, “Mini”)
25g, 21g, 19g
Temperature Range
(1)
5 - 45 degrees Celsius
EMG Contact Dimensions
5 x 1 mm
Contact Material
(2)
99.99% silver
1)
Exposure beyond these temperature limits may damage the rechargeable battery.
2) Sensor skin contacts are made from pure silver and should not be used if allergic reactions to silver are expected or found to occur.
Electrical Specifications
RF Frequency Band
2400-2483 MHz (ISM band)
EMG Signal Input Range
11mV / 22mV r.t.i.
EMG Signal Bandwidth
20-450 Hz / 10-850 Hz
Accelerometer Range
±2g, ±4g, ±8g, ±16g
Accelerometer Bandwidth
24 Hz – 246 Hz (configurable in software)l
Gyroscope Range
±250 dps, ±500 dps. ±1000dps, ±2000dps
Gyroscope Bandwidth
24Hz – 361 Hz (configurable in software)
Magnetometer Range
±4900 uT
Magnetometer Bandwidth
50 Hz
Inter-Sensor Delay
< 1 sample period (Base Station only)
Intra-Channel Delay
< 1-2 sample period