Trigno Quattro, Duo, Mini EMG Sensor User’s Guide
13
“Mini” Analog Output Configuration
Sampling Rate
Data Type
Bandwidth
Ch. x.1
1926 sa/sec
EMG
20-450 Hz
Ch. x.2
148 sa/sec
ACC or Gyro X
DC-50 Hz
Ch. x.3
148 sa/sec
ACC or Gyro Y
DC-50 Hz
Ch. x.4
148 sa/sec
ACC or Gyro Z
DC-50 Hz
Table 4: “Mini” Analog Output signal details. Note that sampling rates are approximate;
please refer to specification table for precise sampling periods.
Figure 3: “Mini” Analog Output Data Flowchart
**Refer to the Trigno System User Guide for more information on Analog Output
Operation.**
Placing the Trigno Sensor on the Skin
Trigno Quattro, Duo and Mini Sensors consist of a main sensor body and four,
two or one cabled detection head(s) respectively. The sensor heads should be
placed near the centroid of the muscle to detect the maximum amount of EMG
activity. The sensor body should be affixed in a convenient nearby location to
provide a detection reference point. The sensor body and sensor head are
easily attached to the skin using the Delsys Adhesive Sensor Interfaces. The
sensor arrow on the top side should be oriented parallel to the direction of the
muscle fibers.
Figure 6. The EMG sensor head must be positioned overtop of the muscle of interest. The
reference contacts can be positioned in a convenient nearby location.
EMG contacts
reference contacts