MALFUNCTION
POSSIBLE CAUSES
CORRECTIVE ACTIONS
Red Alarm indicator light
does not illuminate
during alarm conditions.
Alarm indicator light
not working.
Confirm that alarm indicator light illuminates
during self-check upon power-up. Contact Delfi.
No cuff pressure and/or
tourniquet time reading.
Faulty pressure/time
display.
Internal hardware
failure.
Confirm that all segments of the display
illuminate during self-check upon power-up.
Contact Delfi.
Pump runs continuously.
External leak (cuff or
hose).
Internal leak.
Internal hardware
failure.
Correct leak to clear leak alarm.
Test for leaks (see Section 3.4 above).
Disconnect P.T.S. from cuff to deflate cuff if
required. Contact Delfi.
BATT ERROR alarm.
Fully discharged
battery
Battery pack
disconnected.
Faulty or dead battery
pack.
Connect to AC power and allow battery to
recharge for 8 hours.
Remove battery cover and ensure battery pack is
securely plugged in.
Replace battery pack (see Section 3.5 above).
Battery does not charge
when unit is plugged in
to AC power.
Faulty or dead battery
pack.
Replace battery pack (see Section 3.5 above).
Unit does not turn off
(cannot be set to
standby)
Pressure in cuff
Internal hardware
failure
Deflate cuff and ensure it deflates fully
(disconnect hose if required) to clear “CUFF NOT
DEFLATED” alarm.
Contact Delfi. Unit can be powered off by
unplugging the AC power and removing the
battery pack (see Section 3.5).