Condition
Description
Message
display
(steady)
Alarm
tone
Pressure
display
Time
display
Action to silence
alarm tone for 30
seconds
Action required to correct
alarm condition
Leak in cuff
or hose
The unit has detected a
large leak in a cuff
defined as the cuff
failing to reach the
pressure set point in a
reasonable time, or the
pump running
excessively while
regulating in ‘Inflating’
or ‘Regulating’ mode
and the pressure set
point is not being
adjusted.
CUFF LEAK
(may also
display
PRESSURE
LOW)
Constant
Current
cuff
pressure,
flashing
Normal
Momentarily Press
‘PR ’,
‘PRESSURE – ‘,
‘TIME + ‘, or
‘TIME – ‘.
Fix leak
OR
Reduce pressure set point until
unit can maintain pressure
OR
Press and hold ‘DEFLATE’ for 2
seconds to deflate the cuff.
Calibration
failure
The unit has detected
that the calibration in
the pressure transducer
is invalid. The unit has
shut down in a safe
state*.
(See also Calibration
Procedure, Section 3.3)
ERROR
Constant
“CAL”
“Err”
None
Disconnect the hose and press
‘ON/STANDBY’ twice to shut unit
down and restart
OR
Calibrate the unit per Section 3.3
Internal
electronic
failure
The unit has detected
an internal error and
has shut down in a
safe state*.
ERROR
Constant
“Err”
Error code
None
Press ‘ON/STANDBY’ twice to
shut unit down and restart.
When cuff deflation is required,
disconnect hose to cuff.
* In the ‘safe state’ mode, the current pressure in the cuff is held (in the absence of leaks). The ‘safe state’ mode helps prevent unexpected
loss of occlusion during a procedure if a sudden failure occurs. See Section 2.8.2 for more detail.