Condition
Description
Message
display
(steady)
Alarm
tone
Pressure
display
Time
display
Action to silence
alarm tone for 30
seconds
Action required to correct
alarm condition
Attempt to
power off
unit with
cuff press-
urized
The unit has detected
that the operator has
attempted to set the
unit to standby with
pressure remaining in
the cuff.
CUFF NOT
DEFLATED
Constant
while
button
held, and
for 2
seconds
after
button
released.
Current
cuff
pressure,
flashing
Normal
Momentarily Press
‘PR ’,
‘PRESSURE – ‘,
‘TIME + ‘, or
‘TIME – ‘.
Press and hold ‘DEFLATE’ for 2
seconds to deflate the cuff, then
press ‘ON/STANDBY’ to shut unit
down.
Cuff
pressuri-
zed during
power up
The unit has detected
an approximate
pressure of 50mmHg or
more in the cuff when
the unit was powered
up. The unit assumes
a procedure is in
progress, changes the
pressure set point to
the detected pressure,
immediately regulates
the cuff to this
pressure, and activates
alarm to alert the user
of the condition.
PRESSURE
Constant
Current
cuff
pressure,
flashing
Normal
Momentarily Press
‘TIME + ‘ or
‘TIME – ‘.
Momentarily press
‘PR ’ or
‘PRESSURE – ‘ button to display
pressure set point
OR
Adjust pressure set point by
pressing and holding
‘PR ’ or
‘PRESSURE – ‘.
Cuff not
fully
deflated
The unit has detected
pressure in the cuff
greater than 15mmHg,
60 seconds after the
cuff has been deflated.
CUFF NOT
DEFLATED
Constant
Current
cuff
pressure,
flashing
Normal
Momentarily Press
‘PR ’,
‘PRESSURE – ‘,
‘TIME + ‘, or
‘TIME – ‘.
Disconnect hose to cuff to
exhaust pressure.
1
5