Using the sterile tweezers, carefully
remove a sterile membrane filter from
the packet. Hold the membrane only by the
edge and do not let the membrane filter touch
anything while it is being transferred to the
filtration apparatus.
Remove the sterile vacuum cup from the
filtration apparatus. Push the filtration
apparatus firmly onto the vacuum cup (see
Appendix E
for assembling the filtration
manifold). Place the assembly in an upright
position in a convenient place in the kit.
DO NOT place the apparatus on the ground
where it may become soiled.
Unscrew the plastic collar and filtration
funnel in order that these may be easily
removed. DO NOT place these on any surface
other than the filtration base.