Evaluation and Repair of the Kit
49
7.2.2 Procedure for Recalibrating the Incubator
Leave the testing lid and thermometer
assembly in place and keep the incubator
switched on.
Insert the trimmer tool into the hole on
the side of the unit and locate the tool in
the calibration screw (blue trim).
1
2
To increase the temperature, turn
the adjustment screw clockwise.
To decrease the temperature, turn the
adjustment screw anti-clockwise. Make the
adjustments in stages, a ¼ turn at a time.
After each adjustment, leave the incubator to
stabilise for at least 30 minutes. The complete
recalibration procedure may take several
hours.
Once the incubator has been
recalibrated to read between 43.5 and
44.5°C, leave it switched on for at least 3 hours.
Take note of the temperature at 30 minute
intervals to ensure that the temperature is
stable.
The following day, switch on the
incubator and allow it to reach a stable
temperature. If the temperature is not within
the correct limit, repeat the recalibration
process detailed in
STEPS 1
to
4
.
Dismantle the temperature checking
equipment and store in a safe place.
Empty the water out of the incubator and dry
the inner surfaces.
Switch off the incubator and leave to
cool. DO NOT disconnect the incubator
from the mains electricity supply.
3
4
6
7
5
i
Note:
The above procedure
guarantees an average incubator
temperature of 44°C +/- 0.5°C. After
reaching the set temperature, the
temperature in the incubator may vary
/- 0.5°C during incubation.