OXFAM-DELAGUA Water Testing Kit — Users Manual
35
Pour the medium onto the pad
2.
Allow the medium to warm to ambient
temperature before use. Pour enough
culture medium onto the absorbent pad
in the petri dish to soak the pad and
leave a slight excess (approximately
2.5ml). Try not to let the medium run
down the side of the bottle onto the
pad, and do not allow the bottle neck
to come into contact with any external
objects. Replace the bottle cap
immediately. If too much medium
has been added, drain off the excess
and wipe up any spillage with an absorbent cloth or tissue. Always ensure that
a slight excess remains in the petri-dish to prevent the pad drying out during
incubation.
Note: Once the bottle of culture medium has been opened, it is recommended that
the contents are used within one day. It is not advisable to use the medium in one
bottle over several days since this can lead to contamination.
3.
Flame the tips of the tweezers
with the lighter for
approximately 5 seconds and
leave to cool.
Flame the tips of the tweezers