Phantom Installation, Operation & Service Manual
6-16
Release Cable Replacement
3
Remove x-ray head according to Section “X-Ray Head Replacement”
on page 6-8.
4
Remove locking cylinder according to Section “Locking Cylinder
Replacement” on page 6-13.
5
Unscrew four cover screws (1 in Figure 6-18) and remove cover (2).
Figure 6-18. Cover
6
Remove retaining ring (1 in Figure 6-19) and pull pin (2) out of bracket
(3).
Figure 6-19. Retaining Ring and Pin
1 (x4)
2
1
3
2
Содержание DynaRad Phantom-8000
Страница 6: ...Phantom Installation Operation and Service Manual 4...
Страница 24: ...Phantom Installation Operation Service Manual i 18 Applicable Standards...
Страница 26: ...Phantom Installation Operation Service Manual ii 2...
Страница 66: ...Phantom Installation Operation Service Manual 3 24 Folding Phantom...
Страница 74: ...Phantom Installation Operation Service Manual 4 8 Check tube head rotation friction lock...
Страница 84: ...Phantom Installation Operation Service Manual 5 10 Adjustments 5 3 Adjustments Not applicable for this product...
Страница 106: ...Phantom Installation Operation Service Manual 6 22 Release Cable Replacement...
Страница 116: ...Phantom Installation Operation Service Manual 8 2 Electrical Schematics...
Страница 117: ......
Страница 118: ......
Страница 119: ......
Страница 120: ......
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 126: ......
Страница 127: ......
Страница 147: ...Phantom Installation Operation Service Manual 9 20 Control Chassis Assembly 501203...
Страница 148: ...Phantom Installation Operation Service Manual n 1 Notes...
Страница 149: ...Phantom Installation Operation Service Manual n 2 Notes...
Страница 150: ...Phantom Installation Operation Service Manual n 3 Notes...
Страница 151: ...Phantom Installation Operation Service Manual n 4 Notes...