Attaching the ceiling adapter to a wooden ceiling
Use the adjustment gutters provided. Place them (3x) on to the bracket sections. Lift the assembly arm/brackets into
position and attach it to the ceiling with 3x M8x50 DIN 571 screws.
Use M8x40 DIN 916 screws to adjust the horizontality of the assembly. The arm has to be vertical according to either
front/back or left/right axis.
2. Connection of the power supply cables.
The light must be hard wired to the mains (not plug connected). Every connection not enough fixed is able to disturb the
normal operating of the light. Route the wires into the connector box and attach them to their place properly, according
to the local installation requirements.
An external 2-poles mains switch that fulfils the local installation requirements must be provided by installator.
See principle diagram here bellow:
External 2-poles
Mains switch
Min : 2A/250V
Earth
Neutral
Phase
Input mains