PRO-TC-MAN-17V01 (0802)
Copyright © 2017 Define Instruments
17
channel to calibrate. (To enter step 6.6, you must select
SKIP
at 6.5A.)
If calibration fails,
_ _ _ CALIBRATION FAILED
will scroll across the display twice,
and then you will be directed back to 6.5A to try calibrating again.
The most likely cause of this message is that the controller has not detected any change in input
signal during the calibration process. Check your signal and connections, and then repeat the
calibration procedure.
6.6 - Analog output setup
A _ _ _ ANALOG OUTPUT SETUP
scrolls across the display and toggles with
SKIP
.
If your controller does not have analog output installed, (or you do not wish to
configure your analog output now), press
to skip to 6.7.
Otherwise, press the
button and then
to
ENTER
analog output setup.
B _ _ _ DATA SOURCE FOR ANALOG O/P
scrolls across the display and toggles
with the current analog output data source. Use the
and
buttons to
select an option from:
DISP
,
TEMP 1
,
TEMP 2
,
TEMP 3
,
TEMP 4
,
AVETMP
,
MAXTMP
,
MINTMP
,
PEAK
,
VALLEY
or
MPX
(multiplex). Then press
.
Note that your options in this step will be limited by the 'Number of Sensors' selected in 6.2D.
See Section 5 for more information on analog output data sources, with specific reference to
5.4C for the
MPX
(multiplex) feature.
C _ _ _ LOW SCALE VALUE FOR ANALOG O/P
scrolls across the display and tog-
gles with the currently selected low scale display value. Use the
and
but-
tons and
buttons to enter your cal low position, and then press
.
D _ _ _ HIGH SCALE VALUE FOR ANALOG O/P
scrolls across the display and
toggles with the currently selected high scale display value. Use the
and
buttons to enter your cal high position, and then press
.