System Date
The
System date
option allows the user to set the date of the internal AED clock.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
System date
:
AED Main Menu
Ú
AED options
Ú
System date
What it does
: The
System date
option allows the user to set the date of internal AED clock. Once
the
System date
option is selected, press the CENTER softkey button to enter set date mode:
The year selection will be highlighted green:
• Press the TOP or BOTTOM softkey buttons to adjust the year.
• Press the CENTER softkey button to accept the year setting.
The month selection will be highlighted green:
• Press the TOP or BOTTOM softkey buttons to adjust the month.
• Press the CENTER softkey button to accept the month setting.
The day selection will be highlighted green:
• Press the TOP or BOTTOM softkey buttons to adjust the day.
• Press the CENTER softkey button to accept the day setting.
The date is now set and the user may use TOP or BOTTOM softkey buttons to navigate to other choices
in the menu.
Note
: The factory default setting of the internal AED clock is Universal Time (GMT).
Volume Level
The
Volume level
option allows the user to set AED audio output to
high
,
medium
, or
low
volume.
Changing the volume level will not change the volume of the Active Status Indicator “beep.”
Before you begin
: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
Volume level
:
AED Main Menu
Ú
AED options
Ú
Volume level
What it does
: The
Volume level
option allows the user to set AED audio to
high
,
medium
or
low
volume. Once the
Volume level
option is selected, use the TOP and BOTTOM softkey buttons to
cycle through the different volume settings. When the desired volume selection is chosen, press the
CENTER softkey button to set that volume level. The AED will now use that volume setting for all audio
(except the volume of the Active Status Indicator “beep”). The user may use the TOP or BOTTOM
softkey buttons to navigate to other choices in the menu.
Note
: The factory default setting of the volume level is “
High
.”
Audio Recording
The
Audio recording
option allows the user to enable or disable recording of event audio data to a
Defibtech Data Card (DDC card).
Before you begin
: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
Audio recording
:
AED Main Menu
Ú
AED options
Ú
Audio recording
What it does
: The
Audio recording
option allows the user to enable/disable recording of event audio
data. Once the
Audio recording
option item is selected, use the TOP and BOTTOM softkey buttons
to select either enabled or disabled settings. When the desired selection is chosen, press the CENTER
softkey button to set the feature. The AED will now use that audio recording setting. The user may use
the TOP or BOTTOM softkey buttons to navigate to other choices in the menu.
Note
: The factory default setting of the audio recording is “
Disabled
.”
6. Maintenance Mode
43
DAC-U2510EN-BF rev H